OZONE QC Check FORM
Date: 08/08/06 Time: 8 AM Site: Tribal Air Office QC Check Performed By: SD
Standard Certification: Certified by Tribe QC Standard Source: BGI DeltaCal External Standard?: No
Site O3 Sampler
Make: / Dasibi / Sample Freq:Model: / 1008-AH / Control Freq:
Serial Number: / IN001 / Span:
Full Scale: / 500 / Sample Flow: / 4.8 LPM
Zero Device Display: / 28 PPB / Pass Zero Check? / L1Z , no
Shelter Temperature: / 22 oC / Temp Within Range? / --- , yes
Shelter Pressure: / 760 mm Hg
Ozone QC Device
Make: / Dasibi / Ozone Lamp Temp.: / oCModel: / 1008-PC / Phot. Lamp Temp.: / oC
Serial Number: / IN007 / Phot. Pressure / mm Hg
Det IO: / Phot. Temp.: / oC
Det I:
QC Concentration Check (Values in Units of PPB )
QC Standard Known Value / Analyzer Value / DAS Value / % Difference / % Difference Within Range?55 / 54 / 54 / 1.82 / --- , yes
90 / 90.1 / 90.2 / 0.22 / --- , yes
180 / 180.8 / 180.8 / 0.44 / --- , yes
400 / 402.4 / 402.4 / 0.60 / --- , yes
Linear Regression Chart: y = mx + b
Analyzer DAS
m / 1.0089 / 1.0088b / -1.0363 / -0.9884
r2 / 1 / 1
Note: All of the data in this document were artificially generated for the purpose of this course only and are not from actual instrumentation. Some of the data were generated to purposely produce failing QC check results. The data are NOT representative of actual instrument performance.
OZONE QC Check FORM
Date: 08/22/06 Time: 8 AM Site: Tribal Air Office QC Check Performed By: SD
Standard Certification: Certified by Tribe QC Standard Source: BGI DeltaCal External Standard?: No
Site O3 Sampler
Make: / Dasibi / Sample Freq:Model: / 1008-AH / Control Freq:
Serial Number: / IN001 / Span:
Full Scale: / 500 / Sample Flow: / 4.8 LPM
Zero Device Display: / 2 PPB / Pass Zero Check? / --- , yes
Shelter Temperature: / 24.6 oC / Temp Within Range? / --- , yes
Shelter Pressure: / 760 mm Hg
Ozone QC Device
Make: / Dasibi / Ozone Lamp Temp.: / oCModel: / 1008-PC / Phot. Lamp Temp.: / oC
Serial Number: / IN007 / Phot. Pressure / mm Hg
Det IO: / Phot. Temp.: / oC
Det I:
QC Concentration Check (Values in Units of PPB )
QC Standard Known Value / Analyzer Value / DAS Value / % Difference / % Difference Within Range?55 / 55.4 / 55.4 / 0.73 / --- , yes
90 / 89.8 / 89.8 / 0.22 / --- , yes
180 / 179.2 / 179.2 / 0.44 / --- , yes
400 / 399 / 399 / 0.25 / --- , yes
Linear Regression Chart: y = mx + b
Analyzer DAS
m / 0.9965 / 0.9965b / 0.2283 / 0.2283
r2 / 1 / 1
Note: All of the data in this document were artificially generated for the purpose of this course only and are not from actual instrumentation. Some of the data were generated to purposely produce failing QC check results. The data are NOT representative of actual instrument performance.