General Instructions for Laboratory

General Instructions for Laboratory

PHYSICS 536

GENERAL INSTRUCTIONS FOR LABORATORY

A. INTRODUCTION

A teaching laboratory has special problems. You cannot assume that components and instruments are working correctly. A preceding student may have returned a faulty item to storage without recognizing the problem. Diodes, transistors, integrated circuits, and electrolytic capacitors are usually reliable Always report suspected instrument problems to the instructor.

Learn the limitations and capabilities of the instruments so that you get good results. The equipment is not fragile, but it can be damaged. The following are common mistakes.

  1. Instruments pushed of tables.
  2. Dropped scope probes.
  3. Ignoring voltage and current limitations of instruments. For example, be sure the analog meter is on the correct scale before it is connected to the circuit.
  4. A bright, stationary spot on a scope face will leave a permanent mark.

Since there are a large number of students using the laboratory, everyone must help keep the lab in order.

B. LAB REPORTS.

Lab reports should be brief and concise. Often it is sufficient to write the expected and observed value without comment. Arrange the results clearly to help the grader. Other sources should be identified; for example, tables in the text, manufacturers specifications, general reference values from the lecture notes, etc.

Formal error analysis is not required. However, you need to learn the accuracy of various components so that you know when your results are reasonable. An unreasonable measurement should be recognized in the lab and resolved with the aid of the instructor, if necessary. If the problem is not resolved, it should be described in your report. Graphs, sketches, and tables should be included when requested.

The following material is not required.

  1. Repeat of homework calculation.
  2. Instruction form the experiment description.
  3. Circuit diagrams.
  4. Description of what you did, e.g.”I connected the green wire to the blue terminal, etc.”

You may include materials of this type if you wish, but it will not improve your grade.

C. GENERAL INSTRUCTIONS.

The general instructions given below will be used in many experiments. They are numbered for easy reference in the experiment instructions. These references are essential, and you should not proceed in an experiment without understanding the assigned instructions. Instructions that are not needed in the first lab period are marked with an asterisk. You can study these instructions when they are referenced in subsequent experiments.

1.1 CIRCUIT CONSTRUCTION. Most circuits are assembled using sockets mounted on a metal box. The white sockets running across the short dimension of the board are connected together in sets of five, separated by the center division. Connection errors are the most common problem in the lab.

Be sure you understand the socket arrangement, and check the connections first if the circuit does not work. Arrange the circuit similar to the diagram in the instructions to avoid confusion. This also helps the instructor to find errors if you need assistance. In most experiments, it is not necessary to cut leads if the circuit is neat and safe. However, you may cut leads to improve the arrangement of components and avoid accidental contact between components. Some circuits require special component arrangements, which will be specified. Do not force wire into the plug-in sockets. The wire can go between the metal part of the sockets and the plastic wall, which damages the socket. Try wiggling the wire in the socket until it goes in easily. With practice, you can tell when the wire is going in correctly. You should not insert large diameter wire into the sockets, for example, the meter probes. Connect the meter probe to a component lead of a short wire. Large wire can be forced into the socket, but then it will not make good contact when a normal size wire is used.

1.2 VOLTAGE AND COMMON. There are three sets of colored posts with sockets (red, green, and black). The post and all sockets of the same color are connected together. In addition, the black set is connected to the metal box. The red and green are used to distribute voltages, and the black is used as the circuit common.

1.3 SIGNAL CONNECTORS. Two coaxial connectors are mounted on the box. Each has a wire that can be connected to the white sockets. These connectors are used to connect the circuit to the signal generator and scope.

2.1 METERS. There are several types of meters used in the lab; a hand-held digital multimeter and a benchtop digital multimeter. Familiarize yourself with the various measurement options and ranges for each of these options. The best measurements are made when the appropriate range is selected. These meters are described in more detail in the First Laboratory instructions. The clips on the meter leads should be positioned carefully so that they do not cause connections between components.

2.2* CURRENT MEASUREMENTS. A meter must be in series with a component to measure its current, which usually means breaking into the circuit as shown in Part A of the sketch.

Positive current should go into the plus side of the meter. In the current mode, the meter resistance is low so that the voltage drop across the meter will be small to minimize interference with the circuit. The digital meters can still be used to measure current when it is too small to be observed on the lowest current range. In the voltage mode the meter acts like a 10M resistor, hence a very small current will produce sufficient voltage to be measured. (See Part B of the preceding sketch). For example, a 10-9A current cannot be observed on the most sensitive current ranges, but that current flowing through 10M produces 10mV, which is well within the range of the meters. The large 10M resistance could interfere with the circuit, hence this technique should be used only when the current is small. The current flowing through a resistor R can be determined by measuring the voltage across it (as shown in Part C). When the resistor is large, the internal 10M of the meter must be included in parallel with R to calculate the current.

2.3* VOLTAGE MEASUREMENT. The meter is in parallel with the component for a voltage measurement, hence the meter resistance is very large in the voltage mode. The component and the meter must be treated as a parallel combination, and the meter is negligible if its resistance is much larger than the component resistance. The resistance of the digital meter is 10M. The resistance of the analog meter depends on the voltage range that is used; the resistance is 20K times the full scale voltage. For example, on a 100V scale, the resistance is 10K x 100 = 2M. (The multiplying factor is the full scale voltage, not the observed voltage.)

2.4* AC MEASUREMENTS. In the AC mode a meter measures the effective (or RMS) value of a sine wave. All meters work for 60 cycle, and many meters will work over the whole audio frequency range (20Hz to 20MHz). “True-RMS” meters can be used for nonsinusodial waves, for example to measure the amplitude of noise. Check the meter specifications before you rely on it for anything except 60 cycle sine waves.

3.1 POWER SUPPLY. The power supply has two separate voltage sources in it. The normal limits for each source are 25V and 0.2A. The built-in meter can be used to measure voltage or current for either source. Both sources are protected from excess current. The voltage from the source goes to zero automatically if it is connected to a circuit that draws more than 0.2A. This protection feature can be confusing. You might think that there is something wrong with the source when the voltage drops to zero. However, the fault is most likely in your circuit which draws too much current.

3.2 VOLTAGE SOURCE CONNECTIONS. Each source creates a voltage between a pair of post on the front of the power supply. There is no internal connection between the sources or to ground. The three common connections for the posts are illustrated below. The colors refer to the posts on the circuit board. Black is circuit common.

3.3* USUALLY THE SOURCEshould be adjusted to the correct voltage before it is connected to the circuit. Otherwise a component could be damaged before it was adjusted properly. The voltage sources should be turned-off before the circuit components are changed.

4.1 THE SIGNAL GENERATORcan provide sine or square waves over a wide frequency range (10Hz to 10MHz). The frequency can be read from the generator dial; it is not necessary to measure the frequency on the scope. The amplitude is controlled by the push buttons which are marked with the effective (RMS) value of the sine wave. The peak-to-peak amplitude is 2.8 times the effective value. A continuous amplitude adjustment is included. The amplitude of the signal from the generator should be independent of its frequency. The output resistance of the generator is 50ohms, therefore any load that is comparable to 50 ohms will reduce the signal amplitude when the load is connected.

4.2 THE GENERATOR provides a second signal labeled “1 volt”, which is its effective amplitude. This signal is not affected by the amplitude controls, hence it is useful as a trigger source in some experiment.

4.3 OBSERVE INPUT AND OUTPUT. It is good practice to observe the input and output signal in an experiment. Some unexpected effect at the input might spoil the measurement if you set the input signal and then did not recheck it. Connect the signal from the generator to a “T” connector on channel A of the scope. Then the signal can be carried on to the circuit with a second cable. It is not necessary to have the input signal on the scope display at all times, but it can be observed easily by switching input channels. The T can be disconnected form the scope without affecting the signal going to the circuit if both scope channels are needed to observe signals in the circuit.

4.4* EXTERNAL SOURCE RESISTANCE. The model used to represent a voltage source has a constant voltage applied to a series resistor (rs). We create that pattern in real circuits by connecting the signal generator to the circuit through a real resistor Rs. The amplitude of the signal applied to Rs is adjusted (if necessary) to keep it constant.

5.1 OSCILLOSCOPE. The oscilloscope requires more operator skill than the other lab instruments. You will have ample opportunity to become familiar with a scope during the first lab period. Study the written material, experiment with the controls, and ask questions until you can use the scope correctly.

5.2 INPUT OPTIONS. The vertical inputs have three options; AC, DC, and ground. “AC” places a capacitor in series with the signal. The scope displays changed of the input voltage but is insensitive to the average voltage. This mode is convenient for most sine wave measurements. It must be used to observe small variations of large voltages. The capacitor will produce a DC level shift for pulses (see notes Section 3.04).

5.3* DC INPUT.The “DC” mode uses a direct connection rather than a capacitor in series with the input signal. Therefore the scope displays the total voltage relative to its common line. This mode is convenient when you want to observe variations and the voltage relative to a fixed reference. The input is connected first to the reference, and the vertical position control is adjusted until the trace is on a convenient horizontal line. Then the scope input is connected to the signal. The relationship between the signal and the reference is given by the trace relative to the selected horizontal line. The “ground” mode is provided for convenience when the reference is zero volts. This mode switches the vertical input from the front panel to the scope common and opens the connection to the front panel inside the scope after the vertical position is adjusted. When the zero volt reference is set, the vertical position of the scope can be used as a DC voltage meter.

5.4 PROBES. The signal can be connected to the scope through a probe or a coaxial cable. The probe has the advantage of adding only a little capacitance to the circuit (<10pf), but it attenuates the signal by a factor of ten. The equivalent resistance of the probe is 10M. The attenuation is necessary to reduce the capacitance at the probe tip as explained in text appendix A and in notes 3.10.2. Amplitude measurements would be very inconvenient if the attenuation of the probe depends on the frequency of the signal. The probe is tested by using it to observe a square wave on the scope. (A suitable signal is provided at a front panel connector on the scope.) If the attenuation is independent of frequency, the square wave will not be distorted as shown below. A capacitor in the probe is adjusted to remove the distortion.

If the ground lead of the probe is not connected to the ground of the circuit, a large 60 cycle signal can be observed on the scope. Unfortunately, the ground leads tend to break inside their insulation, which can produce this effect. Check the ground lead if you see 60 cycle on the scope.

5.5* COAXIAL CABLE. When the signal observed is very small, a probe cannot be used because it reduces signal amplitude by a factor of 10. (Non attenuating probe can be purchased, but we do not use them in lab). For small signals, a coaxial cable is used to connect the scope to the circuit through one of the connectors on the metal box. The disadvantage of the cable is that it adds capacitance to the circuit, typically 20pf per foot of cable. Since this capacitance can have several adverse effects, a cable should not be used unless it is essential because of small signal amplitude.

5.6* SINE WAVE AMPLITUDE normally is specified as peak-to-peak, because that is the easiest quantity to measure on the scope. If effective or peak values are used, they will be labeled clearly. Adjust the vertical sensitivity of the scope so that the sine wave covers several centimeters on the display to improve accuracy when the amplitude is measured. When both channels of the scope are used, connect them to one signal initially to insure that they have the same gain. Sine wave amplitude measurements are inconvenient when the peaks are far apart. The horizontal and vertical position must be adjusted to get the peaks close to the grid marks on the scope face. It is more convenient to use a slow horizontal sweet to bring the peaks close together so that there is always some peak close to the grid marks. At very slow speed the signal looks like a continuous band across the scope, which is a convenient display for amplitude measurements. However, you must change the sweet to check the form of the sine wave often enough that you are sure it is not distorted by some fault in the circuit.

5.6A* THE BREAK FREQUENCY is determined by observing the “gain” (vo/vi) as a function of signal frequency. (vi and vo are the amplitude of the input and output signals respectively.) The gain is constant in the “mid-frequency” region, but it decreases by 30% at the “break” frequency. Use a mid-frequency signal to adjust the vertical gains of the scope until the vo amplitude is three divisions and the vi amplitude is 3 divisions (leaving one division to separate the two signals). The fine gain knobs can be adjusted because we are interested in the change in vo/vi, not the actual amplitudes vo and vi. Change the input signal frequency until vo drops from 3 to 2.1 divisions, i.e., by 30% which is the break frequency. If the signal generator is working correctly, the amplitude of vi will be constant when the frequency is changed. Nevertheless, it is good practice to monitor the input signal. (The amplitude of vi will appear to decrease above 1MHz because of the high-frequency attenuation in the model 1222 scope. This attenuation is the same in both scope channels, hence it does not affect the ratio vo/vi. The most convenient way to deal with this scope attenuation is to increase the amplitude of the signal from the generator to maintain the three division display of vi on the scope). When the break frequency measurement is completed, return the vertical fine gain adjustments to the calibrated position, so that the scope is ready for actual (rather than relative) amplitude measurements.

5.6B* PULSE TIME CONSTANTS. You can use the time for a pulse to change by 60% to measure the pulse time constant.

when