Vaughan
Version 2.0 Build 9
Bootloader Version 8
3 November 2016
Summary:
This build is to address issues found during HQA testing. The files in the drop folder are shown below:
· VaughanV2B9_BL10_ W1_0_1.hex - contains the instrument firmware, Bluetooth script and Whisper script
· JtagVaughanV2_0_B09.hex – includes the Bootloader, BLE, and Whisper script and instrument firmware.
Before starting any testing, please clear all logs and ensure the date and time are set correctly.
Documentation:
· Requirements in JAMA – Baseline created 11/03/2016 (Requirements – v2.0 Build 9-> B1 – 11/03/2016)
o SysRS – Instrument
o SysRS – Computing Subsystem
o SysRS – Connectivity Subsystem
o SysRS – Gas Front End (6-series)
o SysRS – HMI Subsystem
o SysRS – Bootloader Subsystem
o SysRs – Integral Sampling Pump
· Code Review: The code review is in process in Crucible under the Vaughan project (VGN-34).
· No Instrument parameters have been added.
The tickets that were addressed in this build 9 are shown below from DEV JIRA2 VAUG project.
Key / SummaryVAUG-1276 / In deep stages of LOBAT several critical functions shut off (FAIL UNSAFE Condition)
VAUG-1274 / CH4/CO2 IR Dual Gas sensor testing with v2.0 build8
VAUG-1269 / Toxic Sensors in Fault Condition during LOBAT testing
VAUG-1275 / Cloud icon is shown with BLE cloud lock set
Vaughan
Version 2.0 Build 8
Bootloader Version 8
26 October 2016
Summary:
This build is to address issues found during HQA testing. The files in the drop folder are shown below:
· VaughanV2B8_BL10_ W1_0_1.hex - contains the instrument firmware, Bluetooth script and Whisper script
· JtagVaughanV2_0_B08.hex – includes the Bootloader, BLE, and Whisper script and instrument firmware.
Before starting any testing, please clear all logs and ensure the date and time are set correctly.
Documentation:
· Requirements in JAMA – Baseline created 10/24/2016 (Requirements – v2.0 Build 7-> B1 – 10/24/2016) - No change from build 7
o SysRS – Instrument
o SysRS – Computing Subsystem
o SysRS – Connectivity Subsystem
o SysRS – Gas Front End (6-series)
o SysRS – HMI Subsystem
o SysRS – Bootloader Subsystem
o SysRs – Integral Sampling Pump
· Code Review: The code review is in process in Crucible under the Vaughan project (VGN-33).
· No Instrument parameters have been added.
The ticket that was addressed in this build is shown below from DEV JIRA2 VAUG project.
Key / SummaryVAUG-1269 / Toxic Sensors in Fault Condition during LOBAT testing
Vaughan
Version 2.0 Build 7
Bootloader Version 8
24 October 2016
Summary:
This build is to address issues reported in JIRA. The files in the drop folder are shown below:
· VaughanV2B7_BL10_ W1_0_1.hex - contains the instrument firmware, Bluetooth script and Whisper script
· JtagVaughanV2_0_B07.hex – includes the Bootloader, BLE, and Whisper script and instrument firmware.
Before starting any testing, please clear all logs and ensure the date and time are set correctly.
Documentation:
· Requirements in JAMA – Baseline created 10/24/2016 (Requirements – v2.0 Build 7-> B1 – 10/24/2016)
o SysRS – Instrument
o SysRS – Computing Subsystem
o SysRS – Connectivity Subsystem
o SysRS – Gas Front End (6-series)
o SysRS – HMI Subsystem
o SysRS – Bootloader Subsystem
o SysRs – Integral Sampling Pump
· Code Review: The code review is in process in Crucible under the Vaughan project (VGN-33).
· No Instrument parameters have been added.
The tickets that were addressed (either fixed or declined) in this build are shown below.
From DEV JIRA2 VPROQA Project:
Key / SummaryVPROQA-591 / iAssign warning not generated (site, user, site/user)
VPROQA-577 / Hendrix cloud data - units field for toxics seems to always be zero
VPROQA-586 / Self Test and Internal Apps issue
VPROQA-588 / Special Firmware: Shortened Datalog/failed accelerometer
From DEV JIRA2 VAUG Project:
Key / SummaryVAUG-1263 / Units in Peer Lost do not have NFC even after silencing alarms
VAUG-1267 / iAssign warnings not working for #1. User and #2. Site. #3. User & Site working properly
VAUG-1264 / Error 504 code was observed twice during LOBAT Wireless Tests
VAUG-1246 / Peer Data Fail screen is not as expected/given in test cases VGN-VTC-737
CCB JIRA VPRO Project:
Key / SummaryVPRO-194 / Change upgrade screen to display "iNet Now" not "Cloud"
VPRO-156 / Change display name to "LEL" for the -P, -S, and -U sensors
VPRO-63 / Shorten the warm-up time on VentisPro when it is docked
Vaughan
Version 2.0 Build 6
Bootloader Version 8
3 October 2016
Summary:
This build is to address issues reported in JIRA. The files in the drop folder are shown below:
· VaughanV2B6_BL9_ W1_0_1.hex - contains the instrument firmware, Bluetooth script and Whisper script
· JtagVaughanV2_0_B06.hex – includes the Bootloader, BLE, and Whisper script and instrument firmware.
Before starting any testing, please clear all logs and ensure the date and time are set correctly.
Documentation:
· Requirements in JAMA – Baseline created 10/03/2016 (Requirements – v2.0 Build 6 -> B1 – 10/03/2016)
o SysRS – Instrument
o SysRS – Computing Subsystem
o SysRS – Connectivity Subsystem
o SysRS – Gas Front End (6-series)
o SysRS – HMI Subsystem
o SysRS – Bootloader Subsystem
o SysRs – Integral Sampling Pump
· Code Review: The code review is in process in Crucible under the Vaughan project.
· No Instrument parameters have been added.
· Instrument Parameter spreadsheet for default values has been updated in WindChill. Please update corresponding values in the internal applications: Vaughan Instrument Parameters v18.xlsx
The tickets that were addressed (either fixed or declined) in this build are shown below.
From DEV JIRA2 VPROQA Project
Key / SummaryVPROQA-562 / Ventis Pros will not show alarms or peers for extra peers in a network greater than 25 instruments
VPROQA-569 / Group lost message disappears from NRS on High/Low alarm clearance
VPROQA-573 / Sensor codes are not sent in Bluetooth messages
VPROQA-563 / Peers over max limit showed random numbers of connected peers
VPROQA-575 / Instrument status byte in bluetooth message showing unexpected values
VPROQA-568 / Instrument does not show group lost message when we take instrument out of range
VPROQA-576 / Only 3 bytes before SN in bluetooth messages
VPROQA-574 / Bytes 93 and 94 framing characters are always 00
VPROQA-506 / Illegal/Unsupported sensors not displayed as error in certain configuration menus
VPROQA-572 / Alarms from two peers can be acknowledged on (Ventis Pro) by pressing Enter button once
From DEV JIRA2 VAUG Project:
Key / SummaryVAUG-1257 / Set Date Screen: "DD" and "YY" are not translated in French, Spanish, and German
VAUG-1254 / Network lost indication appears after 6 minutes instead of 60 seconds
VAUG-1249 / Instrument jumps to next screen when enter button is pressed on Peer List Screen to deselect "Back" option
VAUG-1260 / BCL entries are logged as 3 instead of 1 for O2 dual pair(one O2 is in zero fail)
VAUG-1256 / VPro does not handle network size above 25
VAUG-1261 / Cloud Status not updating
VAUG-1242 / BLE sometimes stops working
VAUG-1244 / The user access code should be settable from 0 to 0xA
VAUG-1243 / If no site is seen and proximity alarms are enabled, the instrument enters proximity alarm.
VAUG-1255 / Instrument does not detect the iAssign beacon "Local" is selected on BLE Service Selection Screen
VAUG-1248 / No Cloud Locked Icon Screen.
VAUG-1246 / Peer Data Fail screen is not as expected/given in test cases VGN-VTC-737
VAUG-1245 / Allow use of a test channel for LENS manufacturing test
VAUG-1266 / Remove register 6018 from the list of registers that have to be set for configuration
VAUG-1251 / VGN-VTC-725 needs to be updated for correct value of register 6021
VAUG-1252 / VGN-VTC-728 needs to be updated for correct value of register 6021
VAUG-1253 / VGN-VTC-733 needs to be updated
VAUG-1250 / VGN-VTC-723 step 4 needs to be updated
CCB JIRA VPRO Project:
Key / SummaryVPRO-195 / Remove iNet Now reference for now
VPRO-137 / Modify Ventis Pro firmware to accommodate behavior of the Dynament sensors 17155304-UM, 17155304-VM, and 17155304-NM so as to allow for a span reserve limit of 100% to be readily achieved
VPRO-62 / In low battery when the instrument is almost in a critical battery failure, user presses PANIC button
VPRO-55 / Unavailable state while downloading bad datalog - Bad datalog created in instrument because sensor was illegal.
Vaughan
Version 2.0 Build 5
Bootloader Version 8
8 September 2016
Summary:
This build is to address issues reported in JIRA. The files in the drop folder are shown below:
· VaughanV2B5_BL8_ W1_0_1.hex - contains the instrument firmware, Bluetooth script and Whisper script
· JtagVaughanV2_0_B05.hex – includes the Bootloader, BLE, and Whisper script and application code
Before starting any testing, please clear all logs and ensure the date and time are set correctly.
For this build, the Beacons are set up to take <30seconds to change the site.
Documentation:
· Requirements in JAMA – Baseline created 09/08/2016 (Requirements – v2.0 Build 5 -> B1 – 09/08/2016)
o SysRS – Instrument
o SysRS – Computing Subsystem
o SysRS – Connectivity Subsystem
o SysRS – Gas Front End (6-series)
o SysRS – HMI Subsystem
o SysRS – Bootloader Subsystem
o SysRs – Integral Sampling Pump
· Code Review: The code review is in process in Crucible under the Vaughan project.
· No Instrument parameters have been added, so instruments upgraded will not need to be configured.
The tickets that were addressed in this build are shown below.
DEV JIRA2: VPROSQA:
Key / SummaryVPROQA-540 / H2S going into OR and CO with PPM of ~600
VPROQA-539 / Watch dog reset During binding
VPROQA-544 / Changing security level of NFC tag through iAssign app makes the tag invalid
VPROQA-549 / Hendrix icon when device is not connected to cloud
VPROQA-499 / iNet instrument upgrade failed (during configuration)
VPROQA-530 / Calibration Failure on the DSX-i
VPROQA-548 / Cloud messages not being parsed correctly
VPROQA-552 / Ventis pro instrument leaving a group and trying to bind to a Safecore
VPROQA-534 / Requirement update for BLE to Cloud upgraded, Restart to Enable screen
VPROQA-536 / NFC Tag cloud feature unlock
VPROQA-546 / Number of hops bounded limit
VPROQA-541 / Unable to save ‘Site’ information using NFC Tag.
VPROQA-554 / NRS of DUT does not show peer low/high alarm icon and instrument serial number during low battery warning
VPROQA-556 / Ventis Pro instrument always shows 0 gas reading for the 6th Sensor on Peer Gas screen
VPROQA-557 / Ventis Pro instrument shows the 6th sensor on Peer Gas screen to be in Low Alarm even if that sensor is in High, STEL or TWA alarm
VPROQA-559 / If instrument is taken into Pump Fault while it is in Mandown Alarm, clearing the Pump Fault also clears the Mandown Alarm
VPROQA-564 / TWA/STEL Alarm indication is the same as a low alarm/high alarm on the NRS and wireless list screen
VPROQA-532 / Pressing Mode/enter button to activate backlight in configuration menu advances screen
VPROQA-535 / ‘Leave Group’ option is available on Wireless screen while no group is joined.
VPROQA-537 / Peer Alarm not acknowledged by the DUT
VPROQA-531 / Enable/Disable Confirmation tone not in Admin menu Sub flow
VPROQA-551 / Top bar of instrument does not cycle with 5 second interval during peer event
VPROQA-558 / Ventis Pro shows Low Alarm Icon for the for the 6th Sensor on Peer Gas Screen instead of Warming Up Icon
VPROQA-560 / Attention High Indicator Pattern generated for Last Teammate Left and Network Lost Warning is different in both the cases
VPROQA-561 / On disabling Peer Alarm Annunciations through Modbus, instrument backlight is also disabled
VPROQA-550 / Positive confirmation tone not generated on successful joining
VPROQA-543 / Pump restarting screen is not displayed after startup pump test is passed
DEV JIRA 2: VAUG Project:
Key / SummaryVAUG-1239 / OOM alarm not working properly
VAUG-1238 / VGN-VTC-739 needs to be updated
VAUG-1237 / Partial Alarm Annunciation is being observed for Peer Panic Alarm on peer instrument.
VAUG-1236 / VGN-VTC-723 need to be updated
VAUG-1235 / In Maintenance sub Menu - Enable Disable BLE Screen content changed
VAUG-1234 / On "Company Information Screen" , an extra 0 with a comma is zero is shown after User Name.
VAUG-1233 / When group lost alarm is enabled, user should have the 'leave group' choice on wireless screen.
VAUG-1232 / Change the meaning of the user access code (1 is most restrictive and 10 is least restrictive). Also, do not show the ', and value' on the user/site screen unless the proximity alarms are enabled
VAUG-1230 / Group (or self) lost time out should be settable between 5 and 30 minutes per VGN-REQ_CON-192
VAUG-1228 / Make the CH4 dual range IR illegal
VAUG-1226 / Investigate a whisper radio turning off
VAUG-1224 / A random issue of screen appearing to be horizontally flipped (upside down). It is observed two times on our side with two different instruments.
VAUG-1223 / In dual sense mode when zeroing (calibration in case of O2) is performed on O2 sensor, the screen shows calibration icon instead of zero icon in top section of screen
VAUG-1222 / Test case needs to be updated to add a new screen in navigation of Alarms submenu
VAUG-1221 / Test Case need to be updated to match latest requirements of Wireless screen
VAUG-1219 / BLE not sending sensor in alarm message
VAUG-1218 / Instrument would not bind or connect
VAUG-1213 / Cal Overdue and Bump Overdue full screen is not displayed. Only overdue icon flashes on NRS
VAUG-1211 / NRS showed a blank peer alarm message at top of screen
VAUG-1209 / During whisper testing one instrument had a watchdog reset
VAUG-1189 / NFC Performance with Gain Reduction Settings
From CCB JIRA:
Key / SummaryVPRO-195 / Remove iNet Now reference for now
VPRO-187 / iAssign user/site reminder alert bug
VPRO-186 / iAssign tag shows user name of "5" instead of <blank> when tag tapped for a second time
VPRO-185 / Update config mode 3.1 title
VPRO-184 / Update title on configuration mode screen 1.13
VPRO-181 / Temperature readings off for diffusion instruments
VPRO-175 / Ventis Pro turning on in chargers after power outage
VPRO-172 / Display user's access level on instrument display
VPRO-171 / Update Proximity Alarm screen
VPRO-170 / Change Proximity Alarm tone and flash
VPRO-167 / Update "cloud" to "iNet Now" on configuration mode screen 1.13
VPRO-164 / Update header on peer list screen in operation mode
VPRO-148 / Change exit behavior on wireless screen in Operation Mode
VPRO-146 / Update config 7.1, 7.2, 7.3, and 7.4 headers from "Wireless" to "LENS Wireless"
VPRO-142 / VPRO-80
Time out from peer readings after 30 seconds
VPRO-141 / VPRO-80
Turn on backlight during peer alarm
VPRO-140 / VPRO-80
Update peer reading bar on home screen
VPRO-137 / Modify Ventis Pro firmware to accommodate behavior of the Dynament sensors 17155304-UM, 17155304-VM, and 17155304-NM so as to allow for a span reserve limit of 100% to be readily achieved
VPRO-113 / Add optional prompt at startup to tap iAssign tag
VPRO-81 / Enable BLE to use iAssign Beacons
VPRO-63 / Shorten the warm-up time on VentisPro when it is docked
Vaughan