HOSPITAL STORE

E.S.I. HOSPITAL BASAIDARAPUR

RING ROAD NEW DELHI 110015

Phone: 25936932

Equipment Purchased in year 2006

Sl.No- / Name of the Equipment / Cost(inclusive of tax)
1- / Hydrolic Fracture Table / Rs- 225000.
2- / Nebulizer / Rs- 20000.
3- / Motorized Bed / Rs- 125000.
4- / Nasal Endoscope / Rs- 123760.
5- / High Pressure Autoclave / Rs- 198000.
6- / Photo Therapy Unit / Rs- 90000.
7- / Cardiac Monitor with defibrillator / Rs- 134000.
8- / Elisa Reader with Washer / Rs- 525000.
9- / Autoclave Vertical (Small) / Rs- 20500.
10- / Donor Couch / Rs 156000.
11- / Anesthesia Work Station / Rs-3200000.
12- / Electro Nerve Stimulator / Rs- 35000.
13- / Fiber Optic Bronchoscope / Rs- 478000.
14- / Automated Coaglometer / Rs- 338000.
15- / Non Invasive Ventilator / Rs- 460000.
16- / X-Ray Process Tank (Manual) / Rs-368000.
17- / Patient Warming System / Rs- 359840.

Specification of Autoclaves Vertical (Small)

1)Electrically operated ( 220 V- 240 V)

2)Vertical type

3)Should be double walled

A)Inner wall made up of s.s.

B)Outer wall made up of m.s / s.s.

4 ) Size depth 550mm x 350 mm ( approx )

5 ) Lid should be made up of stainless steel & provided with tightening device .

6)Should have pressure gauge, water level indicator , steam release valve & safety valve & drain for emptying the Autoclave .

7)Joint less Gasket .

8)Working pressure 10psi to 20psi .

9)To be supplied with cord, plug & stainless steel basket .

10)Should have pedal lifting arrangement to lift the lid or handle to lift the lid .

SPECIFICATION OF ELECTRO NEURO NERVE STIMULATOR

1)Compact device with clear digital display, large screen for easy reading.

2)Auto indicator with several tones for audio detection of the intensity delivered via needle.

3)Choice of setting of intensity- both with rotating button and also digital keys in the side.

4)Predefined settings of pulse duration of 50micro second, 150 micro second and 300 micro second.

5)Pulse frequency option of 1Hz, 2Hz and 4Hz.

6)Precision to the value of 0.2mA for current below 0.5 mA, and above0.5mA, INCREMENT OF 0.1mA also the option of nc/mA.

7)Battery operated with 9V battery.

8)Set should contain cables to be connected to the needle and for earthing.

9)Imported brand.

10)Length 20 cm.

11)Dimensions L-200MM

Width max: 93mm min: 57mm

Height max: 40mm min: 23mm

NERVE STIMULATOR NEEDLE FOR REGIONAL BLOCKS (accessory)

1)Ergonomic hub that is easy to manipulate.

2)300 and 170 bevel needles.

3)50cm removable extension line, composed of three layers.(PE/VYG2/PVC)

4)60cm female electrical connection cable-single use.

5)Should be available in G-23,21,20.

SPECIFICATION OF NON INVASIVE VENILATOR

Should have latest art of technology to provide Non Inversive ventilator.

1)Ventilator should have the following performance date:-

a)mode of ventilator – CPAP, bilevel PCV

b)PEEP - 3-20 mbar

c) Frequency - 8-50bpm

d) Ratio - 1:3 to 2:1

e)PinSp/ pbi – level -5-50 nmbar

f)Trigger - Senstive, Normal

g)Ti Apnoea - 5-60 sec

2)Ventilator should have rescue mode Ventilation in case of pressure sensor failure.

3)Ventilator should have Volume guarantee Ventilation.

4)Ventilator should have built in LCDgraphically display for the monitoring of following parameter :-

a)Real time waveform display of PrVst and Flow Vst.

b)Volume, frequency & PIP.

5)Ventilator should have built in alarm system for the follwing parameters-

a)Supply pressure low

b)Airway Pressure low/high

c)Leakage, Apnoea & frequency

d)Low battery.

6)Ventilator should be operatable on mains and battery with back up of 2 hrs.

7) Ventilator should have facility for oxygen enrichment.

8)Ventilator should be light weight.

9)Ventilator should have low noise level.

10) Ventilator should supply with following accessories:

a)- Breathing Hose

b)– Nasal Mask – Adult & child

SPECIFICATION OF CARDIAC MONITOR with- DEFIBRILLATER

1)Should be able to function as Defibrillater cum cardiac monitor.

2)Energy selection should be possible from 5 to 360 J.

3)Synchronization with R. Wave should be possible.

4)ECG monitor should have multiple gain selection.

5)Should have a built in battery of adequate voltage and amperage. Preferrably , it should be easy to replace .

6)Provision for Multiple ECG leads monitoring eg. I, II, III, aVR, aVL, aVF, & Chest leads should be available.

7) All leads should be defibrillation and electrosurgery protected.

8)The monitor display should be of good resolution.

9)The defibrillator – Monitor should be light – weight, portable and user friendly.

10)Should be ragged and work satisfactorily during voltage fluctuation ( or else a compatible a voltage stabilizer should be provided.)

11)A wheeled trolley should be provided as an accessory.

12)Standard accessories to be stated in the literature

SPECIFICATION OF NASAL ENDOSCOPE

1) Nasal Endoscope: - Wide angle, auto clavable, rigid 30 degree/25 degree forward obligue telescope, 4 mm in diameter using Hopkins rod lens system and incorporated Quantity

Source: - Imported /Indigenous

(2) Cold light fountain Source: -

a) Cold light foundation with a minimum 150 watts Halogen lamp with twin outlets to fit standard fiberoptic cable, built in spare lamp, adjustable light intensity.

(b) Standard Fiberoptic Cable with an adapter to fit in outlet of cold light source, 3.5mm and minimum length 180 cm.

(c) Light weight, adjustable head band with cold light reflector having suitable universal adaptor for attaching having suitable universal adaptor for attaching fiberoptic Cable.

Source: - Imported/ Indigenous

.

SPECIFICATION OF DONOR COUCH.

Following specifications were fixed for donor couch.

(1.) Comfortable providing excellent spiral support to the donor.

(2.) Ability to adjust body position automatically through electric motor.

(3.) Shifting of position easy, smooth and noiseless.

(4.) Electric Control for shifting of position easily accessible and available on both ends of couch.

(5.) Facility to raise both ends manually.

(6.) Arm bar assembly easy to adjust with swinging out as well as moving up and down facility.

(7.) Arm rest wide, Sturdy, must hold donor’s arm in comfortable horizontal position throughout blood donation.

(8). Height at the centre of the seat approximately 2 feet including upholstery.

(9.) Lifting Capacity minimum 150 Kg.

(10.) Input power 120v – 240v.

SPECIFICATION FOR HORIZONTAL AUTOCLAVE.

(High Pressure Autoclave)

1) High pressure autoclave 1.2 kg / cm2 i.e. 121 degree C.

2.) Inner chamber stainless steel 316 grade.

3. Outer Chamber stainless steel 304 grade.

4.) Jacket of best quality steel.

5.) Chamber diameter 400mm.

6.) Chamber depth 600mm.

7.) Steam trap.

8.) Steam generator feed water pump.

9.) Self locking safety door.

10.) Tight sealing door gasket (Gasket of heat resistant material silicon /EPDM).

11.) Drain temp. gauge.

12.) Vaccum breakers.

13.) Safety valve for jacket.

14.) Pressure switches, pressure gauge.

15. Low water protection system (Automatic).

16.) Digital temp. indicator.

17.) Micro processor based automatic system.

18.) Should carry ISI/BIS mark (IS 3829).

19.) Provision for vaccum drying.

20.) Power requirement 440 volts (Three phase).

21.) Should be mounted on a stand which is resistant to corrosion.

22.) Should have a report of hydraulic testing.

23. Working principle-Downward displacement of air.

24. Approx. Cost Rs.3.00 Lacs. (each)

Qty.-2 No.

SPECIFICATION OF FIBER-OPTIC BRONCHOSCOPE

1)Flexible Fibroptic Scope with suction channel.

2)Optical system should have at least 110 degree to 120 degree field view.

3)Distal end should have outer diameter of 5.1mm.

4)Insertion tube have outer diameter of 5.2mm.

5)Total length – at least 820mm-1220mm.

6)Working length -600mm.

7)Upward bending 180 degree.

8)Downward bending130 degree.

9)Insertion tube should be marked with bands.

10)Suction channel should have inner diameter not less than 2.6mm.

11)Should have telescopic eye piece compatible to any make of CCD camera.

12)It should have light source of 150 Watt halogen light with inbuilt air pump for automatic

leakage tester.

13)Light source should be of same company and compatible with scope.

14)Leakage tester – Automatic leakage testing device.

15)Should be simplified ,jointless channel.

16)Mouth guard to keep the mouth open during oral intubation.

SPECIFICATION OF X- RAY PROCESSING TANK

1)Manual Film processing tank should have following parts with specification that follows and build

to work smoothly and process about 30- 35 films of all sizes per hours and easy repair for a long

time. It should be able to fit in the space provided without alteration.

Master tank should be made of 16 guage sheet S.S., grade 316, having MAT finish rustproof with integrated heating & cooling arrangements & safe light , illuminator hook and stirrer etc. Tank

should rest on stainless steel support and have a ground clearance of 6 ½” , by rust proof legs

made of 10G steel. Tank top should be made of single sheet stainless ,band to provide front

margins raised by 1 ½ “ and back margins as vertical / slanted or combined 4” and without

any joint with easy access to valves and outlets for repair . Any connections from the bottom

should be minimized to drains from master tank and washing tank for cleaning.

a) Master tank should be designed to fit in a developer & fixer tank of 45 liters and rinser tank of 20 liters capacity with separate lid for develop a fixer tank with separate handles attached to developer & fixer tanks for lifting & cleaning from master tank. Master tank should be insulated with puff or equivalent suitable material. Rinser and washing tank should have a running flow of 3 & 10 liters per minutes from roof top tank.

b) Heating element should not come in direct contact with water such as soil immersion sealed unit element and should be controlled by thermostat temp range from 25-40 degree C with a sensitivity of +/- 1 degree display. Thermostat probe should be placed in a water proof rust proof compartment.

c)Cooling should be done by a separate water cooler unit made of stainless steel with tank capacity of 100 liters or more. It should feed the master tank from the top with regulator valve & provided with filter for dust/ soil particlesfrom clogging the cooler pipes. It should be provided with automatic voltage stabilizer.

d)All inlets should be controlled by inlet gate valve preferably from sides with good quality flexible connectors & should feed tank from bottom from other side of inlet by L bend of stainless pipe.

e)Safe light be illuminator should be of standard quality and should be separately installed.

f)Film hook for retriving fallen films & hangers and stirrer made of good quality S.S.Strip.

SPECIFICATION OF HYDRAULIC FRACTURE TABLE

1)Hydraulic Fracture Table should have hydraulic mechanism for changing the height.

2) It should have Radio translucent top suitable for use with C arm with provision ( like side rail & adjustable clamp for acceptable accessories .

3)The specification of table top.

a)Height – 745mm- 1155mm.

b)Side rail dimension -25 x10mm.

c)Trendlen burg - 27

d) Reverse burg -27

e)Lateral tilting to both sides -18.

f)Head section & Foot section should raised to 30 & in it reverse position to 90 from the trunk position.

g)Base & column fully covered with non- corrosive ( S.S –sheet).

h) Base should be eccentrically placed.

i)The table should have provision of Ortho paedic Traction attachment..

j) It should have X- ray cassette channel & frame set for X- ray cassette.

k)It should have accessories for lower limb & upper limb Traction.

l)Traction attachment should not interfere with movement of C arm.

SPECIFICATION OF AUTOMATED COAGULUMETER

1)Measuring System – Photometric / Denstomatric

2)Automatic Trigger - Sample loading.

3)Automation Termination –Clot Formation.

4)Should perform following tests- PT, APTT, Fibrinogen, TT.

5)Incubator – Built in at 37 degree C .

6)Measuring Cells – Two in No. ( with mixer in built ).

7)Separate position for Reagents and Tests.

8)LCD display & calibration curves.

9)Calculator to show results in seconds , %, INR, g/L Mg/U.

10)Should use unicrovolumes of sample.

11)Keypad control Panel, RS232 part and printer should be there with instrument.

12)Good after sale service ( To bring performance report from uses.)

Indigenous / Imported

SPECIFICATION OF PATIENT WARMING SYSTEM

1)– Should be suitable for intra – operative applications.

2)– Should consist of active warming arm – cum –shoulder section , pair of leg segments and 1 double segments to cover the entire body.

3)– Size Double segment ( 60-62 ) cm x (80-85) cm.

- Arm & Shoulder section (35-40 ) cm x (175-180) cm.

- Leg Segment ( 80-85) cm.

4)–Each double segment & arm cum shoulder segment should have two temperature sensors each for precise temperature control.

5)– Double segment & arm cum shoulder segment should be divided in two sections capable of being switched ON or OFF independently depending upon the nature of surgery and condition of patient .

6)– Should have a control unit to regulate warmth to every area precisely by use of carbon fibers.

7)–Control unit should be capable of warming at least three segments at a time.

8)– Should offer precise digital temperature control with selectable temperature range of 30 to 42 degree C in steps of 0.5 degree C.

9)– Control panel should display intended and actual temperature .

10)– Should have safety features such as Automatic check. Precise temperature control between warming system and patient. Autostop on detecting any problem.

11)– Should have non latex anti- bacterially coated blood and fluid Resistant covers.

12)– Covers should be washable and replaceable.

13)– The control unit should be light weight not more than 2.5 kg small in size (200 x 120 x230 mm approx) and easily attachable to IV rod / OT table with fixing claw.

14)– Should have low energy consumption and noiseless operation.

Specification of Nebulizer

1- Particle Size range - 0.5 to 5u ( Micron)

2- Piston based.

3- Chamber size- 8-10ml.

4- Operation of Nebuligator pressure- 8-10 PSI.

5- Compressor- Litre flow ( Air flow) around 8-10 litre/min.

6- Suitable for Hospital & Emergency settings.

7- Rate of Nebulization- 0.25 to 0.48ml / min.

8- Minimum Noise Label for hospital setting.

Specification of Motorized Bed

1- Approximate Size 2158Lx970Wx450-800H(mm).

2- Retractable back rest & knee rest.

3- Four section perforated Top.

4- Remote should be control back rest, knee rest, back rest & knee rest simultaneously & height.

5- Should have control panel on foot rest.

6- Removable head & foot ends.

7- Should have electrical shock protection.

8- Should have retractable batteries.

9- swing down type side rails.

Specification of Double Surface Photo Therapy Unit

--It should have double surface provide once facility 6-8 MW/OM2/NM at 45cm distance.

-- Ones head surface should have four blue and two white 2f+ Tld fluorescent tubes with individual switching facility.

-- Under surface should have four blue and two white 2f+TLD fluorescent tubes with individual switching facilities.

-- For both the surface white filter should be provided.

--Unit should be fabricated is stainless steel.

--Height adjustment facility should be provided.

--Infant care trolly should be made of transparent acrylic bed and should be attached with under surface unit.

--A baby mattern along with bubble mattern should be provided.

--Drop loan sides should slide dour straight to occupy is space.

-- Head tilting facility should be there.

--Whole unit should be mounted on castor whole with easy mobility and looking device.

-- Time totalizer unit be there.

-- Should be capacity of being used as a single surface phototherapy unit.

--Voltage stabilizer must be provided cuter unit.

--AMC rates after warranty period must be provided for five years with parts and without parts.

Specifications OF ANAESTHESIA WORK STATION

Anaesthesia Workstation integrating Aneasthetic gases flow delivery,

vaporization and monitoring having the following:-

A. ANAESTHESIA MACHINE with following features:-

------

-Anaesthesia Work Surface: Constructed from Welded tubular/

preferably s.s.

-Stainless steel top, shelves and drawers to be provided

-Front wheels having brake/footrest.

-Gas Blocks (regulators) set at 60 psi (having metal

diaphragms)

-Provision for accessory mounting system.

-Oxygen & Nitrous Oxide Yokes and to have non-interchangeable

gas supply inlet- for Oxygen, Nitrous Oxide and Air with color

coded H.P. antistatic tubes with pipeline adapters.

-Separate colour coded large gauges to indicate cylinder and

pipeline pressures.

-Having inbuilt Hypoxic Guard-patient safety chain link

mechanism incorporating nominal basal flow of 200 ml for

low flow anaesthetic techniques, independently of Oxygen

flow control and also to ensure minimum 25% oxygen at

all oxygen and Nitrous Oxide mixtures.

-Having additional small regulators for Oxygen (19 psi) and

Nitrous Oxide(40 psi)to provide stable input to the flowmeters.

-Automatic cut-off of Nitrous Oxide in the absence of Oxygen

supply.

-Having reservoir based audible Oxygen failure alarm of at least

7 secs.

-Dual Cascaded Flowmeter for Oxygen and Nitrous Oxide

accurately calibrated with an accuracy of +/- 2.5% and range:

i.Oxygen- Low flow 200 ml/min.950ml/min

- High flow1 Ltr./min.10 Ltr./min.

ii.Nitrous Oxide-Low flow40 ml/min.950ml/min.

-High flow 1 ltr./min.10 ltr./min.

iii.Air -Single 1 ltr./min.15 ltr./min

-Flowmeter

-Having Oxygen and Air, pneumatic power outlets operating 60 psi to

operate Ventilators.

-Emergency Oxygen flow of at least 35 lit/min. with non-

lockable push button to be provided.

-Having 2 latest vaporizers for Halothane and Isoflurane-

both should be temperature, pressure and flow compensated,

with key filling arrangement and should be quick mountable

and to allow operation of one vaporizer at a time automatically.

These vaporiser should have NON SPILL system to limit movement

of liquid agent if the vaporizer is momentarily tilted or inverted

and maintaining output within clinically acceptable limits.

B.The Circle Absorber(Closed Circuit)

- Large capacity double chamber reversible sodalime canister.

-Transparent canister so that colour change of sodalime is

seen easily.

-Condensation trap at the base of central tube prevents

accumulation of moisture in sodalime.

-Unidirectional inspiratory and expiratory valves in control head

in addition to an adjustable air way pressure limiting valve

-A lever allows canister to be excluded from gas circuit when

not required or during refilling.

-Mechanism for discouraging channeling of gases.

-Sodalime in lower chamber should exhaust first.

-Quick changeover switch from Bag to Ventilator.

AdditionalPort for measuring

C.ANAESTHESIA MONITOR

------

-Should have high resolution CRT display which can show up to

4 waveforms. The size of screen should be at least 9" to 10".

-The monitor should not require any lengthy start-up

procedures or calibration. It should be ready to monitor as

soon as On/Off switch is pressed.

-Optional facility of thermal recorder/plain paper printer.

-Should be able to monitor the following parameters.

-Should have alarms.

(i)ECG

---

-3 or 5 lead ECG cable.

-Waveform frequency response should be from 0.5 to 25 Hz.

-Heart range should be from 30 to 250 beats/min with accuracy