CT FACIAL
INDICATIONS: WithOut:Trauma, fracture, pain
With:Mass, infection
PATIENT PREP: WithOut:No Prep
With:NPO for solid foods, water only, for 4 hours prior to scan
IV CONTRAST: Perform as directed by the supervising radiologist.
WithOut:None
With:80mL Isovue 370 @ 2.0 mL/second.
ORAL CONTRAST: None
POSITIONING: Head First Supine. Position head to superimpose the base of the skull and acanthion.
TOPOGRAMS: AP. Range: Below mandible through frontal sinuses. 35 mA, 120 kV.
Lateral. Range: Below mandible through frontal sinuses. 35 mA, 120 kV.
SCAN TYPE: Spiral/Helical
NOTES: Instruct patient to keep eyes closed and not roll them during scan.
ALARA – Keep radiation dose As Low As Reasonably Achievable.
FACIAL
ScanRange / Scan Direction
Scan Type / Respiration / Scan
Delay
(Seconds) / CARE
DOSE
4D / Quality
Reference mAs / CARE
kV / Quality
Reference
kV / Detector Configuration
Slices Per Tube Rotation / Pitch
Table Increment/Speed:
(mm/rotation) / Rotation
Time
Seconds / SFOV
(cm)
Below Mandible Through
Frontal Sinuses / Caudocranial
Spiral/Helical / Quiet
Respiration / WithOut: 2 Seconds
With:60 Seconds / ON / 125 / Semi / 120 / Detectors: 64 x 0.6 mm
Slices Per Tube Rotation: 32 / Pitch: 0.8
Table Increment/Speed:
15.36 mm/rotation / 1.0
Seconds / 30
cm
Plane / SAFIRE
Strength / Slice
Thickness / Interval / Kernal / Window
Width/Level / DFOV
(cm)
RECON – AXIAL BONE / Axial / OFF / 1.0 mm / 1.0 mm / H60s
Sharp FR / Osteo
1500/450 / FOV just beyond
Patient’s face
RECON – AXIAL SOFT TISSUE / Axial / OFF / 1.0 mm / 1.0 mm / H40s
Medium / Mediastinum
350/50 / FOV just beyond
Patient’s face
RECON – CORONAL BONE MPR
Coronals Angled in Plane to Body Part / Coronals Angled in Plane to Body Part
Recon Card 3D Coronal MPR / OFF / 1.0 mm / 1.0 mm / H60s
Sharp FR / Osteo
1500/450 / FOV just beyond
Patient’s face
RECON – SAGITTAL BONE MPR
Sagittals Angled in Plane to Body Part / Bilateral Straight Sagittals
Angled in Plane to Body Part
Recon Card 3D Sagittal MPR / OFF / 1.0 mm / 1.0 mm / H60s
Sharp FR / Osteo
1500/450 / FOV just beyond
Patient’s face
RECON – CORONAL SOFT TISSUE MPR
Coronals Angled in Plane to Body Part / Coronals Angled in Plane to Body Part
Recon Card 3D Coronal MPR / OFF / 1.0 mm / 1.0 mm / H40s
Medium / Mediastinum
350/50 / FOV just beyond
Patient’s face
RECON – SAGITTAL SOFT TISSUE MPR
Sagittals Angled in Plane to Body Part / Bilateral Straight Sagittals
Angled in Plane to Body Part
Recon Card 3D Sagittal MPR / OFF / 1.0 mm / 1.0 mm / H40s
Medium / Mediastinum
350/50 / FOV just beyond
Patient’s face
*The operator must check the CTDIvol before and after the scan to ensure it is within the allowed dose range. Scans performed outside of the allowed range must be
documented and reviewed by the designated radiologist and/or physicist.
Allowed CTDIvol Dose Ranges:4 mGy – 60 mGy
XR29 Dose Notification Value (CTDIvol):60 mGy
NETWORK:Exam to PACS
12/2017