A)SPECIFICATIONS OF EQUIPMENTS FOR DBT PROJECT WORKS OF THE DEPT. OF PHYSICS

  1. Name of item : System for physical/ magnetic property measurements at low

temperature and high magnetic fields with Vibration Sample

Magnetometer (VSM) and Electrical Transport measurement.

Sl.no / Technical specification
1 / Superconducting magnet / ±9 Tesla with field uniformity ±0.01%
2 / Temperature control / 2K- 400K with accuracy ±1% in zero to full magnetic field.
3 / Vibrating Sample Magnetometer / Fast, sensitive and fully automated vibrating sample magnetometer with longitudinal configuration ( magnetic field, VSM vibration and moment detection all along vertical axis) and VSM oscillation frequency ~40Hz. Accuracy and reproducibility 0.5 %
4 / Helium liquefier / System should have helium liquefier to recover any loss of helium with a minimum liquefaction rate of 10 liters/day of liquid helium or 6 liters/minute of helium gas, whichever is greater.
5 / AC-DC Electrical Transport property measurement / Current range: 10 nA to 500 mA
For certain applications the current may be limited to protect system wiring from damage or self- heating.
Frequency range: DC, 1Hz to 1KHz AC
Resistivity range: AC resistivity 5 wire AC and the capability to do Hall Effect, 1-V curve and critical current measurements in a single add on package.
6 / Computer Peripherals / Operating software with branded P-4 Computer (latest configuration) and printer.
7 / Warranty / System should have a minimum warranty of one year and option for annual maintenance contract.
  1. Name of the item: Atomic Force Microscope to use with Inverted Optical/ NSOM/ Confocal /Fluorescence Microscope

Sl no. / Technical specifications
1 / Scan range / Upto 100 microns.
2 / Scanning mode / Both sample and tip scanning facility.
3 / Type / MFM imaging capabilities.
Liquid cell imaging for biological samples.
Vibration Isolation System.
4 / Compatibility / Compatible with Inverted Optical/NSOM/Confocal /Fluorescence Microscope.
5 / Computer peripherals / Computers, electronics and software for operation.
6 / Warranty / System should have a minimum warranty of one year and should have the option of extending the warranty thereafter.

3. Name of the item: Quantum Mechanics based software for designing nanoclusters

Sl no. / Technical Specifications
1 / Type / Software based on linear scaling density functional method with high accuracy for nanoclusters etc. in materials and life sciences.
2 / Study / The software should be used to study the properties of defects.
3 / Interface / The product should have an interface to the software program.
4 / Product / Should have visualization and statistics software to easy build, visualize, analysis and modeling different material types.
5 / User option / Single user option with perpetual duration.

4. Name of the item: Trinocular Inverted research Microscope with Bright Field,

Dark Field, Phase Contrast and Fluorescence applications

Sl no. / Technical specifications
1 / Type / Trinocular Inverted research Microscope with Bright Field, Dark Field, Phase Contrast and Fluorescence applications with built in illumination. Infinity corrected optical system with Din Standard optics. Automated 5-6 position nose piece and automated universal condenser and necessary phase rings.
2 / Eye piece / 10X / 16 X
3 / Objectives / Fluor Objective (PLAN) Anti Fungus- High grade FLUOR plan objectives designed for Fluorescence microscopy, FLOUR 10X, FLUOR 40X and FLOUR 100X oil immersion.
4 / Focusing / Co axial coarse and fine focusingmechanism with 2-3 step focus controls.
5 / Filters / All basic filters like DLF, Grey Green, Blue etc.
6 / Magnification / 100X-1000X Standard. Optional 1600X.
7 / Imaging system / Latest digital SLR/ CCD camera with live- image capture capabilities, interphase cards, cables and a dedicated imaging software.
8 / Computer / Latest P4 computer with Core 2 Duo processor and graphics card, 19" TFT monitor and printer
9 / Warranty / System should have a minimum warranty of one year post-installation and should have the option of annual maintenance contract thereafter.

5. Name of item: Horizontal Tubular Furnace

Sl.no / Technical specification
1 / Type / Electrically heated horizontal tubular furnace with Programmable PID Controller with phase angle control thyristor drive.
2 / Maximum working temperature / ~ 12000 C
3 / Continuous temperature / ~1100 0C
4 / Accuracy of temperature / ± 10C with digital display
5 / Heating rate / Shall vary through programmable controller with digital display
6 / Tube ID / Approx. 60 mm or more
7 / Tube length / 450 nm or more
8 / Lid required / 2 nos. ( one each on both sides with proper sealing arrangements).

6. Name of item : Drying Oven

Sl.no / Technical specification
1 / Type / Drying Air
3 nos. of racks made of stainless steel
2 / Size / Inside ~height 40"
~Width 24"
~Depth 24"
3 / Temperature range / 1500 C(approx.)
4 / Controller / Thermostatically controlled temperature by digital temperature indicator cum controller
5 / Blower / Turbo blower fan with duct to maintain uniform temperature.

7. Name of item: Ultrasonicator

Sl.no / Technical specification
1 / Type / Basic equipment comprising of generator, ultrasonic converter, horn, tips/ probes, power cables.
2 / Frequency / Output frequency 20 KHz ( minimum or higher)
3 / Controlled / Electronic control of the ultrasonic generator
4 / Processor / Continuous processor or on / off pulsing
5 / Pulse duration / Adjustable pulse duration( on & off time) ranging from 1 sec to at least 1 hr. ( minimum or higher )
6 / Temperature / Temperature monitoring to prevent overheating of the specimen
7 / Duty Cycle / 10-90 %
8 / Timer / Programmable timer
9 / Volume / 1 ml to 1000ml.
10 / Power output / 200 W( minimum or higher )
11 / Tuning / Automatic tuning
12 / Chamber / Soundproof chamber
13 / Probe/ microtips / Different sizes of probes/ microtips to sonicate minimum volume of 1 ml & maximum upto 1000 ml ( 2 or 3 nos. depending upon volume of the probes)
14 / Input / 220 – 240 V.

8. Name of item : Laminar Air Flow Workstation

Sl.no / Technical specification
1 / Size / 4 ft. X 2ft. X 2 ft.
2 / Flow / Horizontal Flow Particle retention 0.2 to 0.3 micron
3 / Air Flow / 90 FPM
4 / Noise level / 65 db(A) ±5 %
5 / Motor / 0.25 HP
6 / Speed / 1440 RPM
7 / Power supply / 220 V
8 / Cabinet / Made of water proof ply with rubber paint. The side panels and front doors are made of polycarbonate. Stainless steel Bench Top. Transparent double folded front door, fluorescence illuminator, power sockets.
8 / Essential attachments / Static Pressure manometer and Germicidal UVlamp.

9.Name of item : Analytical Balance

Sl.no / Technical specification
1 / Type / Electronic analytical balance. With the inbuilt auto calibration, 7 displays, tare keys, stability indicator, multi user selectable filter levels, plus optimized display.
2 / Capacity / 100-200 gm
3 / Readability / Accuracy / 0.1 mg / 0.0001 gm.
4 / Tare range( subtractive ) / 120g
5 / Response time (avg.) / 2 s.
6 / Repeatability / ±0.1 mg / 0.0001 gm
7 / Linearity / ±0.2 mg / 0.0002 gm
8 / Pan ( inner diameter) / 80 mm ( single pan)

10. Name of item:Table top air-cooled Centrifuge

Sl.no / Technical specification
1 / Type / Versatile table top air cooled centrifuge with flexibility of interchangeable rotors for 1.5 to 2 ml tubes, 15 ml and 50 ml falcontubes, MTP platesand blood collection tubes.
2 / Maximum speed / 17000 RPM approx.
3 / Maximum RCF / 30000Xg
4 / Noise level / Noise level < 67 db(A)
5 / Rotor / Rotor for 6X15/50 ml Falcon tubes rpm 7500(approx) and rotor for 30X 1.5/2.0 ml tubes rpm 13,000(approx). All rotors should be automatically recognized.
6 / Lid / Automatic lid opening
7 / Speed/RCF conversion / Yes
8 / Power supply / AC, 220 V.

11. Name of item: Microtiter Plate Reader

Sl.no / Technical specification
1 / Spectral Range / 340-750 nm (approx).
2 / Accuracy / +/- 2%
3 / Precision / CV<0.5 %
4 / Linearity / +/-2%
5 / Filters / 8 slot filter wheel with standard filters of 405, 450 and 620 nm and 570 nm as optional.
6 / Plate Movement / Selectable continuous Mode as well as stepping mode.
8 / Shaking / Linear Shaking with minimum 3 speeds.
9 / Display / High contrast LCD display
10 / Optical System / Quartz Tungsten Halogen Lamp. Instrument should be compliant and certified for in vitro diagnostics.
11 / Onboard software / Single/ dual wavelength endpoint, single wavelength kinetic, two point and multi wavelength measurements through an onboard and /or PC based dedicated software.
12 / Onboard memory / Minimum 60 assay protocols, retained blank & retained curve fits.
14 / Printer facility / Option for attachment for external printer through serial and/or USB mode.
15 / Power source / 200-240V. 50Hz.
17 / Installation / Installation free of cost.
18 / Brochure / Brochure must be attached
20 / Warranty / Warranty of 2 years and an annual maintenance contract for after warranty.

B)SPECIFICATIONS OF DIESEL SMOKE METER FOR THE DEPT. OF MECHNICAL ENGINEERING.

Diesel Smoke Meter:

Make:AVL

Measurement range:Opacity:0 to 100 %

Absorption:0 to 99.99 m -1

Resolution:Opacity:0.1 %

Absorption:0.01 m -1

Accuracy:better then ± 0.1 m -1 of measurement value

Operating Temperature:5˚ C to 50˚ C

Linearity check:50 % of measurements range (automatic)

Measurements length:0.43 mm ± 0.005 mm

Dimension:570 ×1250 × 500 mm

( W × H × D )

Weight:50 Kg

Power Supply:220 V, 50 Hz

11.5 to 36 V DC

OPTIONS

  • RPM / Oil temperature
  • Computerization Software available
  • LCD Remote

FEATURES:

  • It should be designed for all test procedures (free acceleration and steady-speed test)
  • It should comply with ECE regulation 24 and ISO 3173

C)SPECIFICATIONS OF GAUSS METER:

1) Gauss Meter, battery operated reads directly magnetic field upto 25K gauss with hall

probe.

2) Gauss Meter, Battery operated reads directly magnetic field upto 25K gauss, with hall

probe mains operated.

Terms & Conditions:

  1. The rates should be quoted along with specifications and supporting documents (technical details and brochures, etc.).
  1. Quotations should be accompanied by-

i. A call deposit (earnest money) amounting to Rs. 5,000/- pledged in favour of

the Registrar, TezpurUniversity, payable at Tezpur.

  1. Quotation for each category/departments should be given in separate covers.
  1. The rates should be exclusive of taxes and applicable tax % should be clearly indicated.
  1. The University is exempted from Custom & Excise duty.
  1. Terms and Conditions of supply should be clearly indicated in the quotations.
  1. Those who have quoted rate for Diesel Smoke Meter in response to our Quotation notice no.-TU/11-19/Quotation/2007/1748-A,dtd-6/7/09 are requested to submit quotation as per the new specifications; such parties need not submit call deposit again.
  1. The supplier should be in a position to deliver and install the ordered items within the stipulated time mentioned in the supply order.
  1. Items must be delivered at Napaam campus of the University at the cost of the supplier.
  1. The rates should be valid at least for a period of six months.
  1. Lowest quotation will not be sole criteria for qualification/selection.
  1. Proprietary items should be quoted with sole Manufacturer’s/ Distributorship’s certificate.
  1. The University reserves the right to accept or reject any or all the quotations without assigning any reason.
  1. Applications should be forwarded to the Registrar for release of call deposit against unaccepted quotations.
  1. No separate tender paper will be issued from the office; one should only download the specifications from the web site.

Sd/- Registrar

Tezpur University