Configuring questionnaire for in-situ process analysis

Details of end user
Customer name:
Plant type or process type
Contact person:
Address:
Preferred contact language:
Phone:
Fax:
E-mail:
Details of Siemens contact person
Sales office/representative:
Date of inquiry:
Inquiry No:
Name:
Address:
Phone:
Fax:
E-mail:
Result of inquiry evaluation (if measurement feasible technical proposal on page 5ff)
This table to be filled out by PD PA AP SUP staff only!
Project No:
Contact personPD PA AP SUP:
Approval result: / Yes (measurement task feasible with standard LDS6)Yes (measurement task feasible with std SitransSL)Yes (via special application only)No (measurement task not feasible)Partially (limitations on pages 5ff)

LDS6: General remarks

The central unit must be placed in a control room environment or similar. Relative humidity must be below 80% and ambient temperature range between 5 – 45 °C. Ambient temperature at the sensors must be in the range -30 – 70 °C. Purging media should be oil- and dust free. Instrument air or N2 should be provided using a 6 mm outer diameter flexible tube. Steam should be provided using DN12 tubing. For service/installation of the sensors, a free space of 60x60x60 cm must be available around both the transmitter and receiver.

Sitrans SL: General remarks

Relative humidity must be below 100% and ambient temperature at the sensors must be in the range
-20 – 55°C. Purging mediumN2 should be provided using a 6 mm outer diameter flexible tube.
N2should be oil- and dust free with dew point < -10°C. In order to achieve full performance N2 purity must be > 99,7 % and an O2 content of 0,01Vol% is recommended. Condensation on optical surfaces must be avoided.For service/installation of the sensors, a free space of 60x60x60 cm must be available around both the transmitter and receiver.

Attention:

- The LDS6 will provide active (self-supplied) 4-20mA analog output signals

- The Sitrans SL will provide passive (non-self-supplied) 4-20mA analog output signals, for Sitrans SL an external isolating power supply (7,5-30V) for each analog output signal is required, e.g. Sitrans I

LDS6: Explanatory drawing of requested information and essential application information

Analytical information and sensor demands

Channel 1

/

Channel 2

/

Channel 3

Measured Component(s) / O2 (0-600°C)O2 high pressure (0-200°C)O2 (600-1200°C)NH3NH3/H2OHClHCl/H2OHFHF/H2OCOCO/CO2CO2H2OOther: / O2 (0-600°C)O2 high pressure (0-200°C)O2 (600-1200°C)NH3NH3/H2OHClHCl/H2OHFHF/H2OCOCO/CO2CO2H2OOther: / O2 (0-600°C)O2 high pressure (0-200°C)O2 (600-1200°C)NH3NH3/H2OHClHCl/H2OHFHF/H2OCOCO/CO2CO2H2OOther:
Component 1
Measuring Range
Typical concentration
Required repeatability± ppm
Component 2, if applicable
Measuring Range
Typical concentration
Required repeatability ± ppm
Gas Temperature in °C,
min-max, (typical)
Gas pressure in mbar absolute,
min-max, (typical)
Measurement path length in m
Dust load in mg/m3,
min-max, (typical)
Desired response time
Harzardous sensor environment / No EEx zoneATEX zone 0/20ATEX zone 1/21ATEX zone 2/22 / No EEx zoneATEX zone 0/20ATEX zone 1/21ATEX zone 2/22 / No EEx zoneATEX zone 0/20ATEX zone 1/21ATEX zone 2/22
Purging media available on-site
/ Instrument airNitrogenSteamAir blowerNo purgingOther: / Instrument airNitrogenSteamAir blowerNo purgingOther: / Instrument airNitrogenSteamAir blowerNo purgingOther:
(pmin, pmax)
Suitable purging tube materials / Stainless steelHastelloyPolypropyleneOther: / Stainless steelHastelloyPolypropyleneOther: / Stainless steelHastelloyPolypropyleneOther:
Required purging tube length / 400 mm800 mm1200 mm / 400 mm800 mm1200 mm / 400 mm800 mm1200 mm
Flange type / DN65/PN6ANSI 4"/150 lbsEngine lab version (DN65/PN6) / DN65/PN6ANSI 4"/150 lbsEngine lab version (DN65/PN6) / DN65/PN6ANSI 4"/150 lbsEngine lab version (DN65/PN6)
Flow cell for extractive measurement / NoYes, unheated, wall mountingYes, unheated, wall mount, pumpYes, heated, wall mount, pumpYes, heated, carrier, pump / NoYes, unheated, wall mountingYes, unheated, wall mount, pumpYes, heated, wall mount, pumpYes, heated, carrier, pump / NoYes, unheated, wall mountingYes, unheated, wall mount, pumpYes, heated, wall mount, pumpYes, heated, carrier, pump
Hybrid cable length in m / 5 m10 m25 m40 m50 mcustomized length (>50m): / 5 m10 m25 m40 m50 mcustomized length (>50m): / 5 m10 m25 m40 m50 mcustomized length (>50m):
Loop cable length in m / 5 m10 m25 mcustomized length (>25m): / 5 m10 m25 mcustomized length (>25m): / 5 m10 m25 mcustomized length (>25m):
User documentation in / GermanEnglishFrenchSpanishItalian
Additional Remarks

Sitrans SL: Explanatory drawing of requested information and essential application information

Analytical information and sensor demands

Measured Component(s) / O2 (0,9..1,1bara;..600°C)O2 (0,7..5bara;..200°C)CO (CO (0,8..1,2bara;..700°C))CO (0,7..2bara;..300°C)Other:
Measuring Range
Typical concentration
Required repeatability ± ppm
Gas Temperature in °C,
min-max, (typical)
Gas pressure in mbar absolute,
min-max, (typical)
Measurement path length in m
Dust load in mg/m3,
min-max, (typical)
Desired response time
Harzardous sensor environment / No Ex zoneATEX zone 1/21ATEX zone 2/22FM Cl I,II,III DIV I T6
Purging media available on-site
/ NitrogenInstrument airOther:
(pmin, pmax)
Suitable purging tube materials / Stainless steelHastelloyOther:
Required purging tube length / 340mmOther:
Flange type / DN50/PN25DN50/PN40ANSI 4"/150 lbs
Loop cable length in m / 5 m10 m25 m
Communication interface / Analog 4-20mAPROFIBUS DPModbus
User documentation in / GermanEnglishFrenchSpanishItalian
Additional Remarks

LDS6 and Sitrans SL: Please specify the major gas components and flow speed of the process gas

Process gas composition (major species only)
Channel 1 / Channel 2 /
Channel 3
Nitrogen / vol%
Min-max, (typical)
Water vapor / vol%
Min-max, (typical)
Oxygen / vol%
Min-max, (typical)
CO2 / vol%
Min-max, (typical)
CO / ppm
Min-max, (typical)
/ vol%
Min-max, (typical)
/ vol%
Min-max, (typical)
/ vol%
Min-max, (typical)
/ vol%
Min-max, (typical)
/ vol%
Min-max, (typical)
Gas speed in m/s
Min-max, (typical)

LDS6: Return Form -to be filled out by PD PA AP SUP staff only!

Specifications, only to be filled out if feasible parameter range differ from those in section 1 or 2.

In case of missing data the catalogue specs are valid.

Channel 1

/

Channel 2

/

Channel 3

Gas Component 1, feasible / YesNoLimited: / YesNoLimited: / YesNoLimited:
Measuring range, ppm / Vol%
Detection Limit, ppm / Vol%
Repeatability, ± ppm / Vol%
Accuracy, ± ppm / Vol%
Gas Component 2, feasible / YesNoLimited: / YesNoLimited: / YesNoLimited:
Measuring range, ppm / Vol%
Detection Limit, ppm / Vol%
Repeatability, ± ppm / Vol%
Accuracy, ± ppm / Vol%
Min. Response time in s
Gas Temperature in °C,
min-max, (typical)
Compensation signal required / YesNo (if variable) (if not variable) / YesNo (if variable) (if not variable) / YesNo (if variable) (if not variable)
Gas pressure in mbar absolute,
min-max, (typical)
Compensation signal required / YesNo (if variable) (if not variable) / YesNo (if variable) (if not variable) / YesNo (if variable) (if not variable)
Measurement path length in m
Dust load in mg/m3,
min-max, (typical)
Harzardous sensor environment / No EEx zoneATEX zone 0/20ATEX zone 1/21ATEX zone 2/22 / No EEx zoneATEX zone 0/20ATEX zone 1/21ATEX zone 2/22 / No EEx zoneATEX zone 0/20ATEX zone 1/21ATEX zone 2/22
Purging process side
/ Instrument airAir blowerSteamNitrogenNo purgingOther: / Instrument airAir blowerSteamNitrogenNo purgingOther: / Instrument airAir blowerSteamNitrogenNo purgingOther:
Purging flow recommended / moderateelevatedhighno purging / moderateelevatedhighno purging / moderateelevatedhighno purging
Purging sensor side / No purgingNitrogenInstrument airOther: / No purgingNitrogenInstrument airOther: / No purgingNitrogenInstrument airOther:
Remarks

Order information

Order No

LDS 6 Central unit,
CD Sensor, ch 1
CD Sensor, ch 2,
CD Sensor, ch 3
Options: central unit
sensor(s)

Sitrans SL: Return Form -to be filled out by PD PA AP SUPstaff only!

Specifications, only to be filled out if feasible parameter range differ from those in section 1 or 2.

In case of missing data the catalogue specs are valid.

Gas Component 1, feasible / YesNoLimited:
Measuring range, ppm / Vol%
Detection Limit, ppm / Vol%
Repeatability, ± ppm / Vol%
Min. Response time in s
Gas Temperature in °C, min-max, (typical)
Compensation signal required / YesNo (if variable) (if not variable)
Gas pressure in mbar absolute,min-max, (typical)
Compensation signal required / YesNo (if variable) (if not variable)
Measurement path length in m
Dust load in mg/m3,min-max, (typical)
Harzardous sensor environment / No Ex zoneATEX zone 1/21ATEX zone 2/22FM Cl I,II,III DIV I T6
Purging process side
/ N2Instrument airno purgingOther:
Purging flow recommended / moderateno purging
Purging sensor side / NitrogenInstrument airno purgingOther:
Remarks

Order information

Order No

Sitrans SL
Options:

Siemens AG, PD PA AP SUP 3 – 76187 Karlsruhe – Germany

Support Request URL: