CT TMJ
CLOSED MOUTH AND OPEN MOUTH
INDICATIONS: TMJ pain, popping, locking, dysfunction
PATIENT PREP: No Prep
IV CONTRAST: None
ORAL CONTRAST: None
POSITIONING: Head First Supine. Position head to superimpose the base of the skull and acanthion.
TOPOGRAMS: AP. Range: Above TMJ through mandibular notch/superior ramus. 25 mA, 110 kV.
Lateral. Range: Above TMJ through mandibular notch/superior ramus. 25 mA, 110 kV.
SCAN TYPE: Spiral/Helical
NOTES: Scan Bilateral TMJs.
Perform Closed Mouth scan And Then Perform Open Mouth scan.
ALARA – Keep radiation dose As Low As Reasonably Achievable.
TMJ
CLOSED MOUTH AND OPEN MOUTH
ScanRange / Scan Direction
Scan Type / Respiration / Scan
Delay
(Seconds) / CARE
DOSE
4D / Effective
mAs / kV / Detector Configuration
Slices Per Tube Rotation / Pitch
Table Increment/Speed: (mm/rotation) / Rotation
Time
(Seconds) / SFOV
(cm)
Above TMJ Through
Mandibular Notch/
Superior Ramus / Craniocaudal
Spiral/Helical / Quiet
Respiration / 3
Seconds / OFF / 70 / 110 / Detectors: 16 x 0.6 mm
Slices Per Tube Rotation: 16 / Pitch: 0.8
Table Increment/Speed: 7.6 mm/rotation / 0.6
Seconds / 30
cm
Plane / Slice
Thickness / Interval / Kernal / Window
Width/Level / DFOV
(cm)
RECON – AXIAL BONE
Axial Bilateral TMJs / Axial Bilateral TMJs
Bone / 0.75 mm / 0.7 mm / H70s
Sharp FR / Osteo
1500/450 / FOV just beyond
Patient’s face
RECON – AXIAL SOFT TISSUE
Axial Bilateral TMJs / Axial Bilateral TMJs
Soft Tissue / 0.75 mm / 0.7 mm / H40s
Medium / Mediastinum
400/40 / FOV just beyond
Patient’s face
RECON – CORONAL BONE MPR
Bilateral Coronals - Straight
Straight – Not Oblique / Bilateral Straight Coronals
Straight – Not Oblique
Recon Card 3D Coronal / 0.75 mm / 0.7 mm / H70s
Sharp FR / Osteo
1500/450 / FOV just beyond
Patient’s face
RECON – SAGITTAL BONE MPR
Bilateral Sagittals – Straight
Straight – Not Oblique / Bilateral Straight Sagittals
Straight - Not Oblique
Recon Card 3D Sagittal / 0.75 mm / 0.7 mm / H70s
Sharp FR / Osteo
1500/450 / FOV just beyond
Patient’s face
*The operator must check the CTDIvol before and after the scan to ensure it is within the allowed dose range. Scans performed outside of the allowed range must be
documented and reviewed by the designated radiologist and/or physicist.
Allowed CTDIvol Dose Ranges: 2 mGy – 60 mGy
XR29 Dose Notification Value (CTDIvol): 60 mGy
NETWORK: Exam to PACS
12/2017