LOT 2. MEDICAL EQUIPMENT FOR CLINIC INVESTIGATIONS
Item / SpecificationsFlow Cytometry System – 1 unit / Flow Cytometry System
Equipment should not be focused on the use of diagnostic tests from only one manufacturer
The winner company must provide training for the end user personnel.
Technical Specification :
The measurement of particles in analyzed suspension (Analyzer) must have at least 3 separate lasers for excitation, coupled to optical micro-cell made from fiber glass.
Wavelengths of excitement to be in the range of standard 488nm, 633nm, 405nm, at nominal power in the range 10-50mW
Focus flow excitement alignment is guaranteed, does not need separate realignment of the frequency.
Fluorescence emission spectrum to be covered in at least 8 regions (corresponding to a wide range of commercially available fluorocromes label samples). Minimum fluorescence range must be typically for FITC, PE, PerCP/PerCP-Cy5.5, PE-Cy7, APC, APC-Cy7; AmCyan, Pacific Blue.
Constructing optical module for collecting light to provide high sensitivity (high flow efficiency of capture) for the fluorescence detectors, filters and reflective components to be replaced easily by the user.
Must have level and pressure sensors for the hydraulic components (liquid-sheath and residues), and have separate compartments for storing fluids.
Automatic initialization procedure (fluid) and cleaning / decontamination.
To guarantee a maximum speed of acquisition of approximately
10000 events/second and a capacity of processing signals of about
60 000 events/second.
To allow automatic compensation of signals both in real time as well as by acquisition, including in relation to other lasers
To allow later upgrades and use the carousel type module and for the 94 wells module.
The dedicated software should run on a compatible computer with performance (graphics, speed of calculation)
Minimum PC requirements
• Pentium® Intel® 4 1.7 GHz processor
(minimum specification)
• System Bus 400 MHz
• 256MB RAM
• 30GB Hard Drive
• Dual Display Video Card
• 700MB RW CD-ROM (40X)
• On-board SCSI Controller
• 4 USB ports, 2 USB controllers
• Integrated sound and Ethernet 10/100
• 17” flat screen display (1280 x 1024 resolution)
Possibility to transfer and store data over the local network.
Color printer for printing the data analysis and interpretations.
Windows based system with MS Office package included.
The software must allow complete acquisition and processing of digital measured signals, and allows full automation device startup and acquisition
Software must manage the standardization of parameters of the components of measurement for quality control of inter-operator.
Software can run independently and work offline, including ensuring data processing by re-allocating the compensation for the data already stored
Configuring the operating programs to be easy, intuitive type, during acquisition to get access to functions and fast setting.
Ensure the function of self-gating (automatic demarcation coordinates on the contour of grouped populations)
Ensure calibration of all measurement parameters by running the reference microbes
Approved to work in the IVD and CE
Must include starting and demonstration packages
Training must be performed by the winning company free of charge.
Water distiller –
1 unit / - Fully automatic water still
- Productivity: 12 l/h;
- Suitable for bench and wall mounting;
- Distillate quality: 2,3 µ-Siemens/cm;
- Stainless steel evaporator;
- Stainless steel condenser;
- Stainless steel storage tank with capacity of min. 20 liters;
- Automatic switch off when tank full, low water, impurities detected;
- Distillate withdrawal through drain cock with continuous and touch position.
- CE mark
Coagulometer with
4 channels – 1 unit / - Principle of measurement - optico-mechanical system
- Microprocessor controlled system
- Automatic measurements
- Simple operating interface with 3 buttons
- Error alarms
- Memory for calibration curves and 8 methodic.
- Automatic data processing by the microprocessor, automatic recalculation of coagulation time in %, ratio, G / L, mg / dl
- 4 alpha-numeric displays 8 bits for each channel
- Sample – blood with Na citrate
- Sample volume – 100mkl per test
- Reagent volume – 200mkl per test
- Built-in thermostat 37°C, 8 positions for samples, 4 positions for reagents, 4 positions for measuring.
- Possibility to connect an external printer(option)
- Productivity: prothrombin time, thrombin – 120 tests per hour, PTT – 60 tests per hour; fibrinogen – 40 tests per hour.
- Possibility to make following tests - PTT, Fibrinogen, Prothrombin time, Thrombin, Factor VIII, Factor IX ,Factor XIII, monomer fibrin complexes (RFMK), and ABP, anti-thrombin III, protein C, ReaPrC / FV, aRFMK, XIIa dependent fibrinolysis, plasminogen, a- 2-antiplazmin
Automatic Biochemistry
System -
1 unit / THROUGHPUT
- Up to 360 tests per hour;
- Up to 450 tests per hour with ISE unit.
REAGENT SYSTEM
- Two rotors with each 24 positions for 25 ml bottles and 8 positions for 5 ml bottles.
All positions can be assigned as R1, R2 and R3.
Adapters for 5 ml bottles in 25 ml positions;
-10 pairs of 25 ml positions can be used for 50 ml bottles;
- Reagent 1 volume 110 - 400 μl;
- Reagent 2 volume 0 - 180 μl;
- Reagent 3 volume 0 - 180 μl;
- Reagent disk compartment is cooled to approx. 12 C below ambient temperature;
- Preheated reagent needles with level detection and integrated mixer;
- Typical reagent consumption 250 μl per test.
- ISE integrated system for K, Na, Cl and CO2 analysis
- Dry electrode systems for K , Na, Cl included.
SAMPLE SYSTEM
- Sample rotor containing:
- 80 barcode read samples positions;
- Inner ring for 20 calibrators and 10 controls;
- Stat and pediatric functionality;
- Continuous loading;
- Internal barcode reading;
- Primary tubes (13 or 16 mm OD);
- All positions can contain 5 ml or 10 ml primary tubes or sample cups;
- Sample volume 1 - 30 μl per test, programmable in steps of 0.1 μl;
- Sample probe with level detection and integrated mixer.
SAMPLE PREDILUTION
(MULTIPLE REAGENT MODE ONLY)
- Pre-programmed ratios selectable from 1:2 to 1:100 with 3 diluents;
- Automatic parent calibrator dilution.
PIPETTING SYSTEM (2)
- Hamilton syringes and valve block;
- Reagent syringe 1000 μl;
- Sample syringe 100 μl.
CUVETTE ROTORS (2)
- Semi-disposable rotor with 48 cuvettes. Path length 7 mm;
> 10.000 tests per rotor;
- Min. measuring volume 220 μl;
- Measuring temperature 37˚C, controlled by Peltier elements.
WASHING UNITS (2)
- Cuvette-washing with 4 x 500 μl of water. The unit is equipped with liquid sensors. Waste is separated into diluted and concentrated (sample/reagent mixture and optional ISE-unit) waste.
Cuvettes are dried before use.
LIGHT SOURCE (2)
- Quartz-iodine lamp 12V-20W.
OPTICS & WAVELENGTH RANGE
- 2 optical units with each a 8 position filter wheel;
- Automatic wavelength selection by 8-position filter wheel (340, 376, 405, 436, 505, 546, 578 and 620 nm);
- Other wavelengths available on request;
- Half bandwidth 8 to 12 nm.
PHOTOMETRIC RANGE
-0.1 to 3.0 Absorbance.
ANALYTICAL MODES
- Kinetic measurement with linearity check;
- Bichromatic end point measurement with or without bichromatic reagent blank and/or sample blank correction;
- Two point measurement;
- Graphic plot of all measuring points;
- Automatic rerun with sample reduction;
- Non-linear calibration curves.
AMBIENT TEMPERATURE
- 15-32 C
- Maximum humidity 80%.
MEASUREMENT CAPABILITIES (SINGLE REAGENT MODE)
- Reagent Absorbance (bichromatic) before sample addition;
- Kinetic during 7 minutes after sample addition;
- End Point (Bichromatic) 11.5 minutes after sample addition;
- Kinetic can contain two points for two-point measurements.
MEASUREMENT CAPABILITIES (DUAL REAGENT MODE)
- Reagent Absorbance (bichromatic) before sample addition;
- Kinetic 1 for 4.5 minutes after sample addition (can be used as sample blank for Kinetic 2);
- Kinetic 2 for 4 minutes after reagent 2 addition;
- Kinetic 1+2 for 8.5 minutes after sample addition;
- Sample blank (bichromatic) before reagent 2;
- Endpoint (bichromatic) 4.5 minutes after sample addition or 11.5 minutes after sample addition;
- Kinetic 1, Kinetic 2 or Kinetic 1 + 2 can contain a minimum measuring time or two points for two-point measurements.
MEASUREMENT CAPABILITIES (TRIPLE REAGENT MODE)
- Additional timings available for three reagent system.
CALCULATION MODES
- Prozone check for immunology tests;
- Cut-off declaration.
QUALITY CONTROL
- Up to 15 different controls can be defined, 3 per test;
- Westgard rules;
- Levey-Jennings plots.
CERTIFICATES
- CE;
- CB;
- UL.
WEIGHT
- 190 kg.
DIMENSIONS (FLOORSTANDING)
- 117 x 115 x 77 cm (W x H excl. Monitor x D)
To be supplied with PC and Software for applications.
Set of digital electronic pipettes with 8 channels -
2 sets / - Variable volume
- Possibility of autoclavation
- Automatic tip ejector
- Accumulator which lasts for at least 4000 pipettings without charging.
- Different types and speeds of aspirations and pipetting of liquids.
- Metrological verifications mentioned in products passport
- Color marking
- Must have an incorporated number from the producer
- Possibility of repair and exchange of the used spare parts.
- The set must be composed as follows:
- 5-50 mkl (accuracy ±0,25-0,75 mkl or 1,5-5,0 %, CV % - 0,7) 1 unit
- 30-300 mkl ( accuracy ±0,30-3,0 mkl or 1,0-3,0 %, CV % - 0,3) 1 unit
Set of digital
mono-channel pipettes -
2 sets / - Variable volume
- Possibility of autoclavation
- Automatic tip ejector
- Different types and speeds of aspirations and pipetting of liquids.
- Metrological verifications mentioned in products passport
- Color marking
- Must have an incorporated number from the producer
- Possibility of repair and exchange of the used spare parts.
- The set must be composed as follows:
- 1-10 mkl (accuracy ±0,025 - ±0,10 mkl, CV % 0,02 - 2,0) 1 unit
- 2-20 mkl (accuracy ±0,06 - ±0,20 mkl, CV % 0,4 - 2,5) 1 unit - 10-100 mkl ( accuracy ±0,3 - ±0,8 mkl, CV % 0,2 – 1,0) 1 unit
- 20-200 mkl ( accuracy ±0,6 - ±3,0 mkl, CV % 0,3 - 1,5) 1 unit
- 100-1000 mkl (accuracy ±0,5 - ±1,5 mkl, CV % 0,2 - 0,6) 1 unit
Set of electronic digital mono-channel pipettes -
4 sets / - Variable volume
- Possibility of autoclavation
- Automatic tip ejector
- Accumulator which lasts for at least 4000 pipettings without charging.
- Different types and speeds of aspirations and pipetting of liquids.
- Metrological verifications mentioned in products passport
- Color marking
- Must have an incorporated number from the producer
- Possibility of repair and exchange of the used spare parts.
- The set must be composed as follows:
- 10-100 mkl (accuracy ±0,3-0,8 mkl or 0,8-3,0 %, CV % - 0,2) 1 unit
- 100-1000 mkl (accuracy ±3,0-6,0 mkl or 0,6-3,0 %, CV % - 0,2) 1 unit
Mono-channel pipette analogue Stepper -
4 units / - Pipette for repeated dosing volume between 10 - 5000 mkl
- Allows to dose liquids up to 44 times in repeating series without filling tip.
- It is universal and can be used with 7 types of different tips volume from 10 up to 5000 mkl
- At each type of tips its possible to set the pipette to 5 different volumes
Universal laboratory centrifuge with cooling –
2 units / - Angle rotor.
- Possibility of using 32 tubes with dimensions (length 120 mm, diameter 17 mm).
- Maximum speed 4500 rpm and minimum speed of 500 rpm
- Speed change 100 rpm
- Electronic timer setting from 1 to 99 minutes (with step of 1 min)
- Possibility to set an infinite time for centrifugations (∞ - infinity).
- Valve for blocking and unblocking of the lid.
- Possibility of opening the door in urgent cases
- Kinetic energy 8600 Nm
- Class Security - 1
- Noise ≤ 58 dB (A) (depending on rotor used)
- Possibility to change the acceleration speed of rotor and make an forced stop.
- Power consumption not more then 4A
- Type of cooling liquid (R 404A)
- View speed and spin time left in real-time
- Cooling range (-20C ...... + 40C) with step 10C
- Possibility of memorizing user programs (not less than 9 programs)
- Motor with the safety against overheating
- Automatically disconnect in case of imbalance
- Automatically detection of the inserted rotor
- The interior of the centrifuge chamber is made of stainless steel
- Possibility of extraction and cleaning of the rotor chamber.
- Presence of specialized service center.
Laboratory centrifuge –
3 units / - Angle rotor.
- Possibility of using 32 tubes with maximum volume 12 ml
- Maximum speed 6000 rpm and minimum speed of 500 rpm
- Speed change 100 rpm
- Electronic timer setting from 1 to 99 minutes (with step of 1 min)
- Possibility to set an infinite time for centrifugations (∞ - infinity).
- Valve for blocking and unblocking of the lid.
- Possibility of opening the door in urgent cases
- View speed and spin time left in real-time
- Possibility of memorizing user programs (not less than 9 programs)
- Motor with the safety against overheating
- Automatically disconnect in case of imbalance
- Automatically detection of the inserted rotor
- The interior of the centrifuge chamber is made of stainless steel
- Possibility of extraction and cleaning of the rotor chamber.
- Presence of specialized service center.
Universal laboratory centrifuge with cooling for Eppendorf tubes -
2 units / - Angle rotor.
- Possibility of using 30 tubes with dimensions (length 38 mm, diameter 11 mm).
- Maximum speed 14000 rpm and minimum speed of 500 rpm
- Speed change 100 rpm
- Electronic timer setting from 1 to 99 minutes (with step of 1 min)
- Possibility to set an infinite time for centrifugations (∞ - infinity).
- Valve for blocking and unblocking of the lid.
- Possibility of opening the door in urgent cases
- Kinetic energy 8600 Nm
- Class Security - 1
- Noise ≤ 58 dB (A) (depending on rotor used)
- Possibility to change the acceleration speed of rotor and make an forced stop.
- Power consumption not more then 4A
- Type of cooling liquid (R 404A)
- View speed and spin time left in real-time
- Cooling range (-20C ...+ 40C) with step 10C
- Possibility of memorizing user programs (not less than 9 programs)
- Motor with the safety against overheating
- Automatically disconnect in case of imbalance
- Automatically detection of the inserted rotor
- The interior of the centrifuge chamber is made of stainless steel
- Possibility of extraction and cleaning of the rotor chamber.
- Presence of specialized service center.
Vertical laboratory freezer
(-15ºC ..... -30 ºC) –
1 unit / - Temperature range -15ºC -30ºC
- Capacity 300-350 liters
- Must have ambient sensors;
-Refrigeration method Forced-air circulation
- Refrigerator not harmfull
- Automatic defrost system
- Adjustable shelves minimum 4 zinc plated
- Equipped with monitor for temperature visualization
- Possibility to register the temperature from inside chamber
- Compressor ½ horse power
- Noise not more than 56 dB
- Power consumption not more than 8,5 kW
- Electric data 220/230 V 50Hz
- Warranty 36 months
Vertical laboratory freezer
(-50ºC ..... -86 ºC) –
1 unit / - Temperature range -50ºC -86ºC
- Capacity 350-400 liters
- Must have ambient sensors;
- Refrigerator not harmfull
- Equipped with monitor for temperature visualization
- Two compressor system with total 2 horse power
- Noise not more than 65 dB
- Possibility to register the temperature from inside chamber
- Power consumption not more than 18 kW
- Electric data 220/230 V 50Hz
- Warranty 36 months
Thermo-container / - The equipment is endowed with 2 different voltages: 12 v and 24 v so it can be attached to the lighter of the car.
- It is provided with liquid containers which allow you to store a set amount of cold. cold storage takes place automatically during the first 2 or 4 hours of operation with the thermostat in medium position. the refrigerator will then have normal cycling. if the power supply is missing, the accumulated cold will allow you to keep the products for a certain number of hours(with a room temperature of +32°c must have 6 hours range)
- The joint at the lid should ensure a perfect seal
- Hinges for handling
- Outside dimensions 570x345x380 mm
- External/Internal material Polyethylene.
- Insulation material 50 mm
- Storage capacity 26 liters
- Cold life without openings: min 129 h at 43 C WHO test procedure E4/Proc/1
- WHO specifications E4/CB.2
General requirements for the equipment: