Annexure I
Specifications for purchase of Portable High Resolution Spectroradiometer with Instrument controller for Cryospheric Studies for the approval of Technical Evaluation Committee: Indent Ref. No. SAC/HPUR/2017003035 dated 21 Feb. 2017
General Features:
1. Spectroradiometer should be compact, handheld, rugged and will be used for field measurements.
2. Spectroradiometer should measure the spectral radiance (W/m2/sr/nm or equi.) and reflectance of a given Lambertian target along with Irradiance (W/m2/nm or equi.) in visible, near infra-red and short wave infra-red wavelengths.
3. Measured spectral radiance by the Spectroradiometer is required to be NIST/SI traceable, and full calibration report along with uncertainty budget is to be provided.
4. Operation of the system should be with menu driven software and operating console.
Mandatory SpecificationsTechnical Parameters / Specifications
Spectral Range (nm) / 350-2500
Spectral Resolution (nm) / ≤ 3.0 @ 700 nm; ≤ 8.0 @1400 nm / 2100 nm
Spectral sampling (nm) / ≤ 1.5 (300-1000 nm) and ≤ 2.0 (≥ 1000 nm)
Wavelength reproducibility (nm) / ≤ 0.1
Wavelength accuracy (nm) / ≤ 0.5 (Calibration certificate required)
Scanning Time (milliseconds) / ≤ 100
FOV (in degrees) / Standard 25 & 8
Maximum radiance / VNIR 2x Solar, SWIR 10x Solar
Input / Integrated flexible fiber cable of length 3.0 m (tolerance 50 cm) with appropriate optics and pistol grip
Noise Equivalent Radiance (NEdL)
(W/cm2/sr/nm) / (1)UV/VNIR- ≤ 1.00 x 10-9 @ 700 nm
(2) SWIR- ≤ 1.50 x 10-9 @ 1400nm
(3) SWIR- ≤ 2.25 x 10-9 @ 2100nm
Input Power Supply / 230 V AC ±10% @ 50 Hz for running time 4-6 hours in field with Battery charging Provision
Communications Interface / Wired communication: Ethernet/USB
Wireless communication: WiFi/Bluetooth
Weight (kg) / ≤ 6.5 (compact with light weight)
Ambient Temperature Range
(0C) / (1) Operating Temperature 0 to 40 (Insulator/cold weather upgradation is must to operate below zero degree.)
(2) Storage Temperature: -10 or below to 45
Calibrations (NIST/SI traceable) / Wavelength, absolute reflectance, radiance,
Irradiance, along with calibration certificate for both 250 & 80 FOV.
Control console / Laptop (Tough-book type) with high end configuration for acquisition, archival & processing spectrometer data, workable in rugged terrain and harsh weather conditions (such as shock proof, water resistant, anti-vibration, drop resistant) with special casing to handle impact in field, with long hour running battery time, screen 13 inch or more Windows®7 or higher version 64-bit laptop with latest configuration
Software / 1. Standard software for acquisitions, display, archival and processing of collected data
2. post processing, data export to standard format like excel and text format (preferred interface with ENVI® software for expanded capabilities)
3. One copy of operation software and manual in suitable storing device
Accessories / 1. Highly durable ergonomic backpack carrier for simplified field work
2. Wheeled Pelican case for protected shipping and storage
3. Reference (like Spectralon) 12 x 12 inch
4. Wooden case for reference (Spectralon)
5. 250 FOV for standard along with Main Spectroradiometer with Radiometric Calibration
6. 80 FOV optics as input with Radiometric Calibration
7. Full sky irradiance Remote cosine receptor with Radiometric Calibration
8. Direct & diffuse irradiance measurement attachment with Radiometric Calibration
9. Spare battery- 2 sets
10. Tripod for laboratory/field experiments
Manual / Operation and maintenance manual (hard copy as well as soft copy in appropriate storage media)
Installation and Training / installation & training at SAC Ahmedabad
Warranty / Three year full warranty
General Requirements:
- Supplier shall furnish all the details in their quote as asked for technical specifications and shall submit compliance sheet for each specification. Deviations, if any, shall be clearly mentioned in quote. Numerical value shall be furnished in compliance statements, wherever required.
- Supplier shall furnish the list of customers where similar item was supplied by them, and workable examples in Himalayan and Polar region under sub-zero temperature conditions.
- Supplier shall furnish detailed catalogue/brochure of the product quoted along with the quotation and bring out all the technical details of the instrument.
- Supplier should give list of accessories items also.
- Quotation is also sought for periodic calibration at fixed interval for both FOVs separately.
- Authorization certificate is required for agent representative.
Spectroradiometer specifications SAC/HPUR/2017003035 dated 21 Feb 2017