Employee Number: ______
Facility/Dept.: ______
Competency: Temporary Pacemaker Management
Objective:To verify the provider’s ability to identify the basic components of a pacemaker, to understand the pacemaker’s indication(s) for use, and to identify the sensing and stimulation thresholds specific to a given patient
Method of Evaluation Key:
O = ObservationRD = Return DemonstrationT = TestV = Verbal
Competency Criteria / Method of Evaluation / Met / Not MetKnowledge:
Identifies the components of a pacemaker: generator & epicardial pacing wires (unipolar & bipolar)
Explains indications for atrial and ventricular pacing
Describes how to protect, label, and secure the pacer wires:
- Identifies appropriate insulating material that can be used to protect the wires from aberrant external microshocks
- Uses proper aseptic technique to clean the pacer exit site
Describes the four common modes of pacing: DOO, VVI, AAI, DDD
Locates the Pacing Indicator Light (Green) and the Sensing Indicator Light (Blue)
Locates the battery indicator light; verbalizes where back-up batteries are available and when to change the batteries
Explains the principles behind sensing and capturing
Locates the emergency DOO button
Skills:
Adjusting sensing
- Set pacemaker RATE to at least 10ppm below the patient’s intrinsic rate (to ensure non-pacing) UNLESS the patient cannot tolerate this hemodynamically**
- Adjust the atrial or ventricular output to prevent the risk of competitive pacing as follows:
- Atrial: Set A OUTPUT to 0.1 mA.
- Ventricular: Set V OUTPUT to 0.1 mA.
- Decrease Sensitivity: Slowly turn the atrial or ventricular sensing dial counterclockwise(to increase mV value) until the SENSE indicator stops flashing. The PACE indicator will flash continuously.
- Increasing Sensitivity:Slowly turn the atrial or ventricular sensing dial clockwise (to decrease mV value) until the SENSEindicator flashes continuously, and the PACE indicator stops. This is the sensing threshold.
- Set the sensitivity to half the threshold value
- Restore RATE and OUTPUT to the previous value
Adjusting output
- Set the pacemaker rate to at least 10 ppm above the patient’s intrinsic rateUNLESS the patient cannot tolerate an increased heart rate***
- Decrease the output by slowly turning the output dial counterclockwise until ECG shows loss of capture
- Increase the output slowly by turning the output dial clockwise until ECG shows consistent capture (The PACE indicator will flash continuously; the SENSE indicator stops) – this is the stimulation threshold
- Set the output value to twice the stimulation threshold
- Restore rate to previous value
Documentation
Documents vital signs, cardiac rhythm, mode of pacing, pacing rate, A/V output, and A/V sensing per unit policy
Documents a skin assessment at the site of the pacer wires per unit policy
Attitude:
Connects the pacemaker to the patient for a minimum of 24 hours post-operatively and as indicated
Places the pacemaker in a readily visible and accessible for location
Provides the patient and family with pacemaker education
References:
- Medtronic: Dual Chamber Temporary External Pacemaker (Model 5392) – Operator’s Manual:
- Nursing Reference Center Plus: Epicardial Pacing Wires: Caring for and Using
Employee’s Signature Employee NumberDate
Validator’s Signature Employee NumberDate
Clinical Excellence & Staff Development/07.2017