ABB Instrumentation Division Inc.

Warminster, Pa.

Variable Area Flowmeter Calibration Test Report

Sales Order No.:RA78020FW Date :12 OCT 2000

Serial No.:00W051300 Line Item:1

Original S/N:609B458U01

(AS RECEIVED)

Meter Size :1 inch Model No.:10A4555

Tube :FP-1-27-G-10 Float :GSVGT

Max Flow :.0213 LBS/SEC SP.GR.:1

Pressure :14.7 PSIA Temperature :80³F

Inscription :LBS/SEC AIR MET @ 14.7 PSIA & 80³F

Gas :AIR Dens(g/cm3) :0.0011997 Z stp : 1.0000 Visc.(CPS) :0.01839

Deno(g/cm3) :0.0011775 Z opt : 1.0000

Calibration Data

Correction factor:.066777 Additional factor:1

Float Den: 8.020 Calibration fluid :AIR

Rig:A-100 Fm 1- 7 Static press.:0 PSIG

Rig:A-10 Fm 8-11 Static press.:.07 PSIG

Viscosity: .01807 cPs Fluid SP GR:1

Z @ Std:1.00000 Z @ opt:1.00000

Barometric pressure:14.618 PSIA Meter Temp.:70.86³F

Test Data Results

Run # Noz. Press. Temp. PSIG % LAST CAL LBS/SEC % MAX

01 F 22.830 67.8 3.4580 100.0 .0213 .02133 -.149

02 F 19.430 67.8 3.7590 90.0 .0192 .01923 -.030

03 F 15.600 67.9 3.7880 80.0 .0170 .01704 -.049

04 E 46.160 68.6 3.8990 70.0 .0150 .01497 +.091

05 E 36.890 68.3 3.7040 60.0 .0128 .01275 +.253

06 E 28.450 68.3 3.7360 50.0 .0107 .01064 +.249

07 E 19.960 68.5 3.5570 40.0 .0086 .00857 +.061

Run # CUBIC FT Seconds PSIG % LAST CAL LBS/SEC % MAX

08 4.0000000 42.0890 3.0020 30.0 .0064 .00643 -.144

09 3.0000000 47.1440 3.0610 20.0 .0043 .00430 -.041

10 2.0000000 44.9580 2.5840 14.0 .0030 .00305 -.077

11 2.0000000 82.0140 2.5830 8.0 .0017 .00167 -.047

The Measurement and Test equipment, used in the calibration of this meter, is traceable to NIST and is controlled by a

Quality System designed to meet the requirements of ANSI/Z540 and ISO 9001.

This Calibration report may not be reproduced, except in full, without written permision.

Test Performed by :G. Slugg

TEST EQUIPMENT

QC # Description

E*938 DIGITEC TIMER

L**16 10 CUBIC FEET BELL

P**35 0-60 PSIG DRUCK

P*417 BAROMETER

P*505 0-15 PSIG RPM

T**31 THERMOMETER

T*190 THERMISTOR

T*191 THERMISTOR

R**32 CRITICAL NOZZLE F

R**33 CRITICAL NOZZLE E

VA REV16, R.A.M. 040500

ABB Instrumentation Division Inc.

Warminster, Pa.

Variable Area Flowmeter Calibration Test Report

Sales Order No.:RA78020FW Date :12 OCT 2000

Serial No.:00W051300 Line Item:1

Original S/N:609B458U01

(AS LEFT)

Meter Size :1 inch Model No.:10A4555

Tube :FP-1-27-G-10 Float :GSVGT

Max Flow :.0213 LBS/SEC SP.GR.:1

Pressure :14.7 PSIA Temperature :80³F

Inscription :LBS/SEC AIR MET @ 14.7 PSIA & 80³F

Gas :AIR Dens(g/cm3) :0.0011997 Z stp : 1.0000 Visc.(CPS) :0.01839

Deno(g/cm3) :0.0011775 Z opt : 1.0000

Calibration Data

Correction factor:.066777 Additional factor:1

Float Den: 8.020 Calibration fluid :AIR

Rig:A-100 Fm 1- 7 Static press.:0 PSIG

Rig:A-10 Fm 8-11 Static press.:.07 PSIG

Viscosity: .01807 cPs Fluid SP GR:1

Z @ Std:1.00000 Z @ opt:1.00000

Barometric pressure:14.618 PSIA Meter Temp.:70.86³F

Test Data Results

Run # Noz. Press. Temp. PSIG % LBS/SEC

01 F 22.830 67.8 3.4580 100.0000 .02133

02 F 19.430 67.8 3.7590 90.0000 .01923

03 F 15.600 67.9 3.7880 80.0000 .01704

04 E 46.160 68.6 3.8990 70.0000 .01497

05 E 36.890 68.3 3.7040 60.0000 .01275

06 E 28.450 68.3 3.7360 50.0000 .01064

07 E 19.960 68.5 3.5570 40.0000 .00857

Run # CUBIC FT Seconds PSIG % LBS/SEC

08 4.0000000 42.0890 3.0020 30.0000 .00643

09 3.0000000 47.1440 3.0610 20.0000 .00430

10 2.0000000 44.9580 2.5840 14.0000 .00305

11 2.0000000 82.0140 2.5830 8.0000 .00167

The Measurement and Test equipment, used in the calibration of this meter, is traceable to NIST and is controlled by a

Quality System designed to meet the requirements of ANSI/Z540 and ISO 9001.

This Calibration report may not be reproduced, except in full, without written permision.

Test Performed by :G. Slugg

TEST EQUIPMENT

QC # Description

E*938 DIGITEC TIMER

L**16 10 CUBIC FEET BELL

P**35 0-60 PSIG DRUCK

P*417 BAROMETER

P*505 0-15 PSIG RPM

T**31 THERMOMETER

T*190 THERMISTOR

T*191 THERMISTOR

R**32 CRITICAL NOZZLE F

R**33 CRITICAL NOZZLE E

VA REV16, R.A.M. 040500