ABB Instrumentation Division Inc.
Warminster, Pa.
Variable Area Flowmeter Calibration Test Report
Sales Order No.:RA78020FW Date :12 OCT 2000
Serial No.:00W051300 Line Item:1
Original S/N:609B458U01
(AS RECEIVED)
Meter Size :1 inch Model No.:10A4555
Tube :FP-1-27-G-10 Float :GSVGT
Max Flow :.0213 LBS/SEC SP.GR.:1
Pressure :14.7 PSIA Temperature :80³F
Inscription :LBS/SEC AIR MET @ 14.7 PSIA & 80³F
Gas :AIR Dens(g/cm3) :0.0011997 Z stp : 1.0000 Visc.(CPS) :0.01839
Deno(g/cm3) :0.0011775 Z opt : 1.0000
Calibration Data
Correction factor:.066777 Additional factor:1
Float Den: 8.020 Calibration fluid :AIR
Rig:A-100 Fm 1- 7 Static press.:0 PSIG
Rig:A-10 Fm 8-11 Static press.:.07 PSIG
Viscosity: .01807 cPs Fluid SP GR:1
Z @ Std:1.00000 Z @ opt:1.00000
Barometric pressure:14.618 PSIA Meter Temp.:70.86³F
Test Data Results
Run # Noz. Press. Temp. PSIG % LAST CAL LBS/SEC % MAX
01 F 22.830 67.8 3.4580 100.0 .0213 .02133 -.149
02 F 19.430 67.8 3.7590 90.0 .0192 .01923 -.030
03 F 15.600 67.9 3.7880 80.0 .0170 .01704 -.049
04 E 46.160 68.6 3.8990 70.0 .0150 .01497 +.091
05 E 36.890 68.3 3.7040 60.0 .0128 .01275 +.253
06 E 28.450 68.3 3.7360 50.0 .0107 .01064 +.249
07 E 19.960 68.5 3.5570 40.0 .0086 .00857 +.061
Run # CUBIC FT Seconds PSIG % LAST CAL LBS/SEC % MAX
08 4.0000000 42.0890 3.0020 30.0 .0064 .00643 -.144
09 3.0000000 47.1440 3.0610 20.0 .0043 .00430 -.041
10 2.0000000 44.9580 2.5840 14.0 .0030 .00305 -.077
11 2.0000000 82.0140 2.5830 8.0 .0017 .00167 -.047
The Measurement and Test equipment, used in the calibration of this meter, is traceable to NIST and is controlled by a
Quality System designed to meet the requirements of ANSI/Z540 and ISO 9001.
This Calibration report may not be reproduced, except in full, without written permision.
Test Performed by :G. Slugg
TEST EQUIPMENT
QC # Description
E*938 DIGITEC TIMER
L**16 10 CUBIC FEET BELL
P**35 0-60 PSIG DRUCK
P*417 BAROMETER
P*505 0-15 PSIG RPM
T**31 THERMOMETER
T*190 THERMISTOR
T*191 THERMISTOR
R**32 CRITICAL NOZZLE F
R**33 CRITICAL NOZZLE E
VA REV16, R.A.M. 040500
ABB Instrumentation Division Inc.
Warminster, Pa.
Variable Area Flowmeter Calibration Test Report
Sales Order No.:RA78020FW Date :12 OCT 2000
Serial No.:00W051300 Line Item:1
Original S/N:609B458U01
(AS LEFT)
Meter Size :1 inch Model No.:10A4555
Tube :FP-1-27-G-10 Float :GSVGT
Max Flow :.0213 LBS/SEC SP.GR.:1
Pressure :14.7 PSIA Temperature :80³F
Inscription :LBS/SEC AIR MET @ 14.7 PSIA & 80³F
Gas :AIR Dens(g/cm3) :0.0011997 Z stp : 1.0000 Visc.(CPS) :0.01839
Deno(g/cm3) :0.0011775 Z opt : 1.0000
Calibration Data
Correction factor:.066777 Additional factor:1
Float Den: 8.020 Calibration fluid :AIR
Rig:A-100 Fm 1- 7 Static press.:0 PSIG
Rig:A-10 Fm 8-11 Static press.:.07 PSIG
Viscosity: .01807 cPs Fluid SP GR:1
Z @ Std:1.00000 Z @ opt:1.00000
Barometric pressure:14.618 PSIA Meter Temp.:70.86³F
Test Data Results
Run # Noz. Press. Temp. PSIG % LBS/SEC
01 F 22.830 67.8 3.4580 100.0000 .02133
02 F 19.430 67.8 3.7590 90.0000 .01923
03 F 15.600 67.9 3.7880 80.0000 .01704
04 E 46.160 68.6 3.8990 70.0000 .01497
05 E 36.890 68.3 3.7040 60.0000 .01275
06 E 28.450 68.3 3.7360 50.0000 .01064
07 E 19.960 68.5 3.5570 40.0000 .00857
Run # CUBIC FT Seconds PSIG % LBS/SEC
08 4.0000000 42.0890 3.0020 30.0000 .00643
09 3.0000000 47.1440 3.0610 20.0000 .00430
10 2.0000000 44.9580 2.5840 14.0000 .00305
11 2.0000000 82.0140 2.5830 8.0000 .00167
The Measurement and Test equipment, used in the calibration of this meter, is traceable to NIST and is controlled by a
Quality System designed to meet the requirements of ANSI/Z540 and ISO 9001.
This Calibration report may not be reproduced, except in full, without written permision.
Test Performed by :G. Slugg
TEST EQUIPMENT
QC # Description
E*938 DIGITEC TIMER
L**16 10 CUBIC FEET BELL
P**35 0-60 PSIG DRUCK
P*417 BAROMETER
P*505 0-15 PSIG RPM
T**31 THERMOMETER
T*190 THERMISTOR
T*191 THERMISTOR
R**32 CRITICAL NOZZLE F
R**33 CRITICAL NOZZLE E
VA REV16, R.A.M. 040500