PS-420/440

User Manual

PS-420 and PS-440 Patient Simulators

P/N 17225

V. 1.80

Copyright ã 2001 by METRON. All rights reserved.

METRON:

USA FRANCE ______NORWAY______

1345 Monroe NW, Suite 255A 30, rue Paul Claudel Travbaneveien 1

Grand Rapids, MI 49505 91000 Evry, France N-7044 Trondheim, Norway

Phone: (+1) 888 863-8766 Phone: (+33) 1 6078 8899 Phone: (+47) 7382 8500

Fax: (+1) 616 454-3350 Fax: (+33) 1 6078 6839 Fax: (+47) 7391 7009

E-mail: E-mail: E-mail:

Disclaimer

METRON provides this publication as is without warranty of any kind, either express or implied, including but not limited to the implied warranties of merchantability or fitness for any particular purpose. Further, METRON reserves the right to revise this publication and to make changes from time to time to the content hereof, without obligation to METRON or its local representatives to notify any person of such revision or changes. Some jurisdictions do not allow disclaimers of expressed or implied warranties in certain transactions; therefore, this statement may not apply to you.

Limited Warranty

METRON warrants that the PS-420/PS-440 Patient Simulator will substantially conform to published specifications and to the documentation, provided that it is used for the purpose for which it was designed. METRON will, for a period of twelve (12) months from date of purchase, replace or repair any defective simulator, if the fault is due to a manufacturing defect. In no event will METRON or its local representatives be liable for direct, indirect, special, incidental, or consequential damages arising out of the use of or inability to use the PS-420/PS-440 Patient Simulator, even if advised of the possibility of such damages. METRON or its local representatives are not responsible for any costs, loss of profits, loss of data, or claims by third parties due to use of, or inability to use the PS-420/PS-440 Patient Simulator. Neither METRON nor its local representatives will accept, nor be bound by any other form of guarantee concerning the PS-420/PS-440 Patient Simulator other than this guarantee. Some jurisdictions do not allow disclaimers of expressed or implied warranties in certain transactions; therefore, this statement may not apply to you.

Trademarks

Duracell is a registered trademark of Duracell, Inc.


Table of Contents

1. Introduction 1-1

1.1 Features 1-1

1.2 Specifications 1-1

1.3 General Information 1-3

2. Installation 2-1

2.1 Receipt, Inspection and Return 2-1

3. Operating PS-420/PS-440 3-1

3.1 Controls and Terminals 3-1

3.2 Power 3-3

3.3 Operating PS-420/440 3-3

4. Simulating Functions 4-1

4.1 Temperature 4-1

4.2 Respiration 4-1

4.3 Blood Pressure 4-3

4.4 ECG/Arrhythmia 4-5

4.5 Cardiac Output 4-9

4.6 ECG Performance Testing 4-10

Appendix A - Codes and Descriptions A-1

A-1 Keypad Control A-1

A-2 Two-Function Keys A-1

A-3 Preset Codes A-1

Appendix B - Temperature Cable Connection Matrices B-1

B-1 PS-420/440 Temperature Cable – YSI 400 Series B-1

B-2 PS-420/440 Temperature Cable – YSI 700 Series B-2

B-3 PS-420/440 Temperature Cable – HP-YSI 400 Series B-3

B-4 PS-420/440 Universal Temperature Cable - Unterminated B-4

Appendix C - Cardiac Output Cable Connection Matrices C-1

C-1 COA-G Cardiac Output Adapter C-1

C-2 COA-HP Cardiac Output Cable – Monitor Type: HP C-2

C-3 COA-1 Cardiac Output Cable – YSI 400 Series C-3

C-4 COA-2 Cardiac Output Cable – YSI 700 Series C-4

Appendix D - Blood Pressure Cable Connection Matrices D-1

D-1 PS-420/440 BP Cable – Monitor Type: GE-2 D-1

D-2 PS-420/440 BP Cable – Monitor Type: GE-3 D-2

D-3 PS-420/440 BP Cable – Monitor Type: HP-1 D-3

D-3 PS-420/440 BP Cable – Monitor Type: HP-2 D-4

D-4 PS-420/440 BP Cable – Monitor Type: HP-3 D-5

D-5 PS-420/440 BP Cable – Monitor Type: HP-4 D-6

D-6 PS-420/440 BP Cable – Monitor Type: HP-8 D-7

D-7 PS-420/440 BP Cable – Monitor Type: HP-9 D-8

D-8 PS-420/440 BP Cable – Monitor Type: KT-1 D-9

D-9 PS-420/440 BP Cable – Monitor Type: LT-1 D-10

D-10 PS-420/440 BP Cable – Monitor Type: MM-1 D-11

D-11 PS-420/440 BP Cable – Monitor Type: MQ-2 D-12

D-12 PS-420/440 BP Cable – Monitor Type: MQ-3 D-13

D-13 PS-420/440 BP Cable – Monitor Type: NK-1 D-14

D-14 PS-420/440 BP Cable – Monitor Type: PB-2 D-15

D-15 PS-420/440 BP Cable – Monitor Type: PC-1 D-16

D-16 PS-420/440 BP Cable – Monitor Type: SL-1 D-17

D-17 PS-420/440 BP Cable – Monitor Type: SM-1 D-18

D-18 PS-420/440 BP Cable – Monitor Type: SM-3 D-19

D-19 PS-420/440 BP Cable – Monitor Type: TK-1 D-20

D-20 PS-420/440 Universal Pressure Cable – Unterminated D-21

Appendix E - PS-420/440 RS-232 Cable Connection Matrix E-1

Appendix F - ERROR REPORT form F-1

Appendix G - IMPROVEMENT suggestion form G-1


Manual Revision Record

This record page is for recording revisions to your PS-420/PS-440 User Manual that have been published by METRON AS or its authorized representatives. We recommend that only the management or facility representative authorized to process changes and revisions to publications:

·  make the pen changes or insert the revised pages;

·  ensure that obsolete pages are withdrawn and either disposed of immediately, or marked as superseded and placed in a superseded document file, and;

·  enter the information below reflecting that the revisions have been entered.

Rev No / Date Entered / Reason / Signature of Person Entering Change
0 / 1999-01-08 / Initial Release / METRON
1 / 1999-06-18 / Addition of appendixes. / METRON
2 / 2000-04-24 / Addition of COA-G instructions. / METRON
1.80-1 / 1-18-01 / General update

This page intentionally left blank.

v

1.  Introduction

This chapter describes the METRON PS-420 and PS-440 Patient Simulators’ features and specifications.

1.1  Features

PS-420 and PS-440 are compact, lightweight high performance simulators designed to be used by trained service technicians for patient monitor testing. Through settings manipulated by the operator they simulate various electrocardiogram, respiration, blood pressure, temperature and cardiac output conditions. The difference between them is in blood pressure simulation: PS-420 offers two-channel simulation, while PS-440 offers four-channel simulation.

1.2  Specifications

1. ECG

12 Leads with independent outputs referenced to RL
Baseline Impedances / 500, 1000, 1500 or 2000 Ohms for Leads I, II and III
High Level Output: / 1000 x (Lead II)
Rates: / 30, 40, 60, 80, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, and 300 BPM
Default Rate: / 80 BPM.
Rate Accuracy: / ± 1% of selection
Adult or Pediatric Waveform
ECG Amplitudes: / 0.5, 1.0, 1.5 and 2.0 mV
Amplitude Accuracy: / ± 2%. (Lead II)
Superimposed Artifact: / 50 and 60 Hz, muscle, baseline wander and respiration
ECG Performance:
Lead II
Square Wave: / 0.125 and 2.0 Hz
Pulse: / 30, 60, and 120 BPM: 60 mS pulse width
Sine Waves: / .5, 5, 10, 40, 50, 60 and 100 Hz (1 mV amplitude, Lead II)
Triangle Wave: / 2.0 and 2.5 Hz
ST Segment Analysis:
Elevated or
Depressed: / -.8 mV to + .8 mV in .1 mV steps


2. Pacemaker

Pacer Spike Amplitude: / 2, 4, 6, 8 and 10 mV in Lead II
Pacer Spike Duration: / 0.1, 0.5, 1.0, 1.5 and 2.0 mS
Asynchronous Pacemaker
Pacer Non-Function
Pacer Non-Capture
Demand Occasional Sinus
Demand Frequent Sinus
AV Sequential

3. Arrhythmia Selections

Based Rate of 80 BPM / PVCs 12 / minute
Sinus Arrhythmia / PVCs 24 / minute
Atrial (PAC) * / Frequent Multifocal PVCs
Missed Beat * / Bigeminy
Atrial Tachycardia / Trigeminy
Atrial Flutter / Pair PVCs *
Nodal (PNC) * / Run 5 PVCs *
Nodal Rhythm / Run 11 PVCs *
Supraventricular Tachycardia / Ventricular Tachycardia
PVC1 Left Ventricular Focus * / Ventricular Fibrillation Coarse/Fine
PVC1 Early, LV Focus * / Asystole
PVC1 R on T, LV Focus * / Conduction Defects
PVC2 Right Ventricular Focus * / First Degree
PVC2 Early, RV Focus * / Second Degree
PVC2 R on T, RV Focus * / Third Degree
Multifocal PVCs * / Right Bundle Branch Block
Atrial Fibrillation Coarse/Fine / Left Bundle Branch Block
PVCs 6 / minute
* Will go to NSR ECG @ 80 BPM after completion

4. Blood Pressure

Input/Output Impedance: / 350 Ohms
Exciter Input Limit: / ± 10 V
Exciter Input Frequency Range: / DC to 4000 Hz
Transducer Sensitivity: / 5 or 40 µV/V/mmHg
Level Accuracy: / ± 1% ± 1 mmHg
Static Levels:
BP 1: / -10, 0, 80, 160, 240, 320 and 400 mmHg
BP 2: / -10, 0, 50, 100, 150, 200 and 240 mmHg
BP 3 (PS-440 Only): / -5, 0, 20, 40, 60, 80 and 100 mmHg
BP 4 (PS-440 Only): / -5, 0, 20, 40, 60, 80 and 100 mmHg
Channel Selections: / PS-420: / PS-440
Arterial 120/80: / Channels 1 and 2 / Channels 1, 2 and 3
Radial Artery 120/80: / Channels 1 and 2 / Channels 1, 2 and 3
Left Ventricle 120/00: / Channels 1 and 2 / Channels 1, 2 and 3
Right Ventricle 25/00: / Channels 1 and 2 / Channels 1, 2 and 3
Central Venous 15/10: / Channel 2 / Channels 2 and 3
Pulmonary Artery 25/10: / Channel 2 / Channels 2 and 3
Pulmonary Wedge 10/2: / Channel 2 / Channels 2 and 3
Left Atrium 14/4: / Channel 2 / Channels 2 and 3
Automatic Swan/Ganz (every 20 seconds): / - / Channel 4
Manual Swan/Ganz (changes every time entry is selected): / - / Channel 4
Synchronized with all normal sinus rates
Physiologically tracks all arrhythmia selections

5.  Cardiac Output Option

Catheter Type: / Baxter-Edwards, 10cc
Blood Temperature: / 37°C / 98.6°F and 36 °C / 95.9 °F
CO for 2 Degrees: / 3, 5, and 7 L/Min
CO for 20 Degrees: / 3, 5, and 7 L/Min
Cal Pulse: / Of 1 degree for 1 second
Of delta 402 Ohms for 4 seconds
Computational Constant
2 Degrees: / 0.561
20 Degrees: / 0.608
Left to Right Shunt / 2 and 20 Degrees
Faulty Injectate / 2 and 20 Degrees
Calibrated or uncalibrated cardiac output waves for 4 different CO values

6.  Respiration

Baseline Impedances: / 500, 1000, 1500, 2000 Ohms (Leads I - III)
Lead Selection: / LL or LA
Impedance Variations: / 0.2, 0.5, 1.0, and 3.0 Ohms
Rates: / 0 (apnea), 15, 20, 30, 40, 60, 80, 100, 120 BPM
Apnea: / 12 sec, 22 sec, 32 sec, and continuous

7.  Temperature

30°C/86°F, 35°C/95°F, 37°C/98.6°F, 40°C/ 104°F, 42°C/107.6°F
Compatible with YSI 400/700 series.

1.3  General Information

Display/Control: / 16-digit alphanumeric numeric display keys
2 switches for Respiratory Leads LL/LA and Power ON/OFF
RS232 Interface: / Bi-directional interface. Baud rate: 9600
ECG Output Connectors / 10 AHA/IEC color-coded connectors accepting ECG snaps and pins.
Power: / 9 V alkaline battery or battery eliminator
Case: / High impact plastic
Weight (w/o batt.): / 0.343 kg / 12.1 oz.
Dimensions: / Height: 3.4 cm / 1.4 in.
Width: 10.7 cm / 4.2 in.
Length: 16.0 cm / 6.3 in.
Part No: / PS-420 Patient Simulator (P.N. 17200)
PS-440 Patient Simulator (P.N. 17250)
Standard Accessories: / PS-420/PS-440 User Manual (P.N. 17225)
High Impact Plastic Carrying Case (P.N. 17260)
Optional Accessories: / 110 V or 220 V AC Adapter/
Battery Eliminator (P.N. 17021)
Cardiac Output Adapter Box,
PS-420/440 (P.N. 17290)
Temperature Cable *
Blood Pressure Cable *
Cardiac Output Cable *
* Contact your local METRON Sales Agent for further details

1-3

2.  Installation

This chapter explains unpacking, receipt inspection and claims, and the general procedures for PS-420/440 setup.

2.1  Receipt, Inspection and Return

1. Inspect the outer box for damage.

2. Carefully unpack all items from the box and check to see that you have the following items:

·  PS-420 Patient Simulator (P.N. 17200); or PS-440 Patient Simulator (P.N. 17250)

·  PS-420/440 User Manual (P.N. 17225)

·  High Impact Plastic Carrying Case (P.N. 17260)

3. If you note physical damage, or if the unit fails to function according to specification, inform the supplier immediately. When METRON or the company’s Sales Agent, is informed, measures will be taken to either repair the unit or dispatch a replacement. The customer will not have to wait for a claim to be investigated by the Sales Agent. The customer should place a new purchase order to ensure delivery.

4. When returning an instrument to METRON, or the Sales Agent, fill out the address label, describe what is wrong with the instrument, and provide the model and serial numbers. If possible, use the original packaging material for return shipping. Otherwise, repack the unit using:

·  a reinforced cardboard box, strong enough to carry the weight of the unit.

·  at least 5 cm of shock-absorbing material around the unit.

·  nonabrasive dust-free material for the other parts.

Repack the unit in a manner to ensure that it cannot shift in the box during shipment.

METRON’s product warranty is on page ii of this manual. The warranty does not cover freight charges. C.O.D. will not be accepted without authorization from METRON or its Sales Agent.


4-11

3.  Operating PS-420/PS-440

This chapter describes the controls and terminals of the PS-420 and PS-440 Patient Simulators. It also provides basic instructions on operating the simulators.

3.1  Controls and Terminals

PS-420/440 Top Panel

1. Battery Eliminator / 2.1 mm micro jack for connecting a 9 V plug-in power supply transformer for use in operating the unit from any standard electrical outlet. The jack polarity is center negative. We recommend you use METRON’s AC Adapter (P.N. 17021), available separately from your METRON Sales Agent.
2. Power Switch / Turns the power on and off.
3. LA - LL Slide Switch / Selects the reference lead, either LA (left arm) or LL (left leg). The position of the switch must correspond to the type of patient monitor in use.

PS-420/440 Front Panel

4. LCD Display / 15 mm x 60 mm (.58 in. x 2.37 in.) window displaying up to two lines of 20 point font.
5. Keypad Control Keys
enter / Enters the selected code line value into memory.
clear / Clears the code line value from the LCD window.
menu scroll / Causes the code line number to increase or decrease. The display arrows indicate which menu scroll to use. The right menu scroll increases the code lines by 1, while the left menu scroll decreases the code lines by 1.
change / These keys are functional when the top line in the LCD window displays the up/down arrows . They allow the user to increment or decrement the current setting. The up change arrow increases the preset codes by 1, while the down change arrow decreases the preset codes by 1.
6. Keypad Soft Keys / These keys have two functions: numeric and one-step selection of simulator functions.
0 / status / -  Enters a numeric 0 to code line.
-  Displays current parameter settings.
1 / temp / -  Enters a numeric 1 to code line.
-  Changes temperature.
2 / resp rate / -  Enters a numeric 2 to code line.
-  Changes respiration rate.
3 / base-line / -  Enters a numeric 3 to code line.
-  Changes baseline resistance.
4 / bp static / -  Enters a numeric 4 to code line.
-  Sets BP channels static levels.
5 / bp wavef. / -  Enters a numeric 5 to code line.
-  Sets BP channels to BP waveforms.
6 / bp zero / -  Enters a numeric 6 to code line.
-  Sets BP channels to zero level.
7 / ecg rate / -  Enters a numeric 7 to code line.
-  Changes ECG rate for NSR.
8 / ecg size / -  Enters a numeric 8 to code line.
-  Changes ECG amplitude (lead II).
9 / nsr / -  Enters a numeric 9 to code line.
-  Selects Normal Sinus Rhythm (80 BPM).
7. ECG Connectors / Ten snap and multi-banana connectors for ECG output, allowing for connection to any twelve-lead ECG. These terminals are labeled and on the top panel. The labels are AHA/IEC color-coded to aid in matching them to corresponding patient leads. Labels and their definitions are as follows:
Label / Definition
RA / R
LA / L
RL / N
LL / F
V1 / C1 to V6 / C6 / Right Arm
Left Arm
Right Leg (Reference or grounded)
Left Leg
V leads (US and Canada). Also referred to as pericardial, precardial or unipolar chest leads, and chest leads (IEC)