Quick Guide: Complete Galileos System Install:
Sidexis XG: 2.3.05 & 2.4; Galileos Server & Workstation: 1.6; SUI: V03_06_02; Implant: 1.6
· (IT Tech) Create PData folder
· Uncheck “simple file sharing” (tools>folder options>view)
· Share with full rights for “everyone” or “Domain users” in both sharing & security tab.
· (IT Tech) Add users to each new workstation if Workgroup; If in Domain, add “cbadmin” w/pwd: “galileos838” to Domain Admins. *Allows Sirona Support admin access to install & repair.
· Insure there are no firewalls, Anti-virus, and/or Window Defender blocking access.
1. SQL Server: One time at “server” or “capture all-in-one workstation”
· Install SQL Express from Sidexis XG disk; Accept defaults
See “Installing SQL” and “ADDING SQL TO EXISTING SERVER” quick-guides if needed.
2. Install Sidexis XG on master station: Usually at capture station.
· Select Multi-station & Central Install
· Select appropriate SQL Server
· Select proper PData location (see above)
· Complete Practice Name Page & Select to register “Later”
· Test adding a patient and opening an image.
* Version: 2.3
· Unzipped NGPIMAN Fix (2.3.05) to PData\Setup folder
· Copy 2 enclosed files to each workstation into C:\Sidexis folder & overwrite file.
3. Install Galaxis 1.6 on RCU “server”: Always on computer supplied with Galileos 3D System.
· (IT Tech) Network conductivity to RCU (IP/Add to Domain or Workgroup/Permissions/Firewall)
· Install Sidexis XG on RCU from PData\Setup folder
· Install Galileos Server 1.6 from Galileos Plug-In Disk. Install RCU Patch 1.6.1
· Insure the created Data Container folder has full rights for “everyone” or “Domain users” in sharing & security tab.
· Put Dongle in RCU
· Uncheck power mgmt in USB properties
· Create “Dongle” folder in PData & paste appropriate file (.xml file from Support Team)
· Install license in Sidexis Manager>Sirona Admin Control License
4. Install Galaxis Workstation on master station
· Install Galaxis workstation from PData\Setup\Galileos folder & Accept defaults
· Add Galileos Unit in Sidexis Manager>SiXABCon
· Check SUI at Software Updates: V03_06_02 (PData\Setup\Galileos\Galileos Devices)
· Check the date on Extended details (if wrong, change by re-entering IP)
· Calibrate Galileos in Service Exposure & Do 2D & 3D Acceptance test
· Take test acquisition
At any time after Sidexis XG & Galaxis Server is installed:
· Create “Implant SW1.6” folder in PDATA\Setup folder & copy Implant 1.6 disk (latest releases of Galileos Implant database can be downloaded from sicat.com)
· Install Sidexis XG & Galileos 1.6 Workstation on workstations (Do not use disks!)
· Install Implant 1.6 where needed from PData\Setup\Implant SW1.6 folder
· Set up buttons and features in Sidexis & Galaxis
· Copy NGSQLBackup folder (PData\Setup\Tools folder) into PData & run on SQL Server to initiate the “SilentBackup.bat” script.
· (IT Tech) add Scheduled Task to SQL Server to run SilentBackup.bat daily.
*Sometimes this will not work on NAS; * Set script to run on SQL host to mapped drive.
· (IT Tech) add PData to Office’s Daily Backup.
· Complete Galileos Passport, calibration and backup forms and send to Galileos Support.
If needed with approval of Sirona Technical Support: DATALINK
· Download Datalink for 3rd party software into a created folder in PData.
http://www.apteryxware.com/datagrabber/downloads/index.shtml
· Install and configure Datalink (delete entries for additional software)
· Do Datalink Update & Sniffer Definitions for appropriate practice software.
· Register the software with all correct client information.
· Install at workstations by creating a shortcut for Datalink.exe from PData\Datalink folder.
* Check “Partners” in Sidexis>Utilities>Configure System>Service
Rev 7/09 RB
Sirona Dental Systems, LLC 4835 Sirona Drive Suite 100 Charlotte, NC 28273
Tel: 1-800-659-5977 Select option 2 & 2 Fax: 1-888-297-8631